inforMED
MalfunctionFPB

MICRO-VOLUME EXTENSION SET

Received May 5, 2021 · Event occurred Jan 31, 2021

Report 1416980-2021-02695 · MDR key 11779654

Device

Generic name

Filter, Infusion Line

Model number

NA

Catalog number

2N3350

Lot number

ASKU

Product problems

  • Crack
  • Disconnection
  • Crack
  • Disconnection

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED A NON-DEHP MICRO-VOLUME EXTENSION SET DISCONNECTED AND APPEARED CRACKED. DURING AN UNSPECIFIED SOLUTION INFUSION VIA A 50ML SYRINGE AT 1ML/HOUR, THE TUBING BECAME DISLODGED DUE TO A ¿POSSIBLE CRACKED CONNECTION PIECE¿. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative

THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. FUNCTIONAL TESTING WAS PERFORMED, INCLUDING PRESSURE AND CLEAR PASSAGE TESTING; AND NO BLOCKAGE OR LEAKS WERE OBSERVED. FURTHERMORE, PULL TESTING WAS COMPLETED WITH NO SEPARATION DETECTED. THE REPORTED CONDITION WAS NOT VERIFIED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.