SITUATE
Received May 4, 2021 · Event occurred Apr 8, 2021
Report 1717344-2021-00657 · MDR key 11768779
Device
Generic name
Counter, Sponge, Surgical
Manufacturer
Covidien Mfg Dc BoulderModel number
01-0031Catalog number
01-0031
Product problems
- False Positive Result
- False Positive Result
Patient
65 YR
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE DEVICE HAD ERRONEOUS DETECTION READ. SPONGE COUNT PERFORMED AT END OF CASE; ALL SPONGES ACCOUNTED FOR. BODY SCAN WITH MAT INDICATED A DETECTION. COUNTS WERE RECONFIRMED TWICE. ALL TABLES AND MAYO STAND WERE MOVED FURTHER AWAY FROM THE SURGICAL FIELD. BODY SCAN WITH MAT CONTINUED TO INDICATE DETECTION. SCAN OF PATIENT WITH WAND COULD NOT DETECT ANY SPONGE. AFTER PATIENT MOVED OFF TABLE AND OUT OF ROOM MULTIPLE MAT SCANS PERFORMED WITH NOTHING ON THE TABLE; ALL GAVE DETECTION. MAT MOVED TO MULTIPLE DIFFERENT ROOMS AND USED WITH MULTIPLE CONSOLES AND CABLES; MAT ALWAYS GIVES DETECTION. THERE WAS NO PATIENT INJURY.
Additional Manufacturer Narrative
H3: EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. THE EVALUATION FOUND NO POTENTIALLY CONTRIBUTING FACTORS, AND THE SAMPLE MET ALL RELATED SPECIFICATIONS. IT WAS REPORTED THAT THE DEVICE GAVE A FALSE POSITIVE DETECTION. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE IDENTIFIED BECAUSE NO RELATED PROBLEM WAS DETECTED WITH THE DEVICE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: B5, E1 (FIRST NAME, LAST NAME, PHONE NUMBER), E2, E3, E4 (EMAIL), G1 (MANUFACTURER NAME, MFR CONTACT FIRST NAME, LAST NAME, EMAIL, PHONE NUMBER), G2, G3. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem
ACCORDING TO THE REPORTER, DURING A MINI FACE/NECK LIFT + BILATERAL ABDOMINOPLASTY PROCEDURE, THE DEVICE HAD ERRONEOUS DETECTION READ. SPONGE COUNT PERFORMED AT END OF CASE; ALL SPONGES ACCOUNTED FOR. BODY SCAN WITH MAT INDICATED A DETECTION. COUNTS WERE RECONFIRMED TWICE. ALL TABLES AND MAYO STAND WERE MOVED FURTHER AWAY FROM THE SURGICAL FIELD. BODY SCAN WITH MAT CONTINUED TO INDICATE DETECTION. SCAN OF PATIENT WITH WAND COULD NOT DETECT ANY SPONGE. AFTER PATIENT MOVED OFF TABLE AND OUT OF ROOM MULTIPLE MAT SCANS PERFORMED WITH NOTHING ON THE TABLE; ALL GAVE DETECTION. MAT MOVED TO MULTIPLE DIFFERENT ROOMS AND USED WITH MULTIPLE CONSOLES AND CABLES; MAT ALWAYS GIVES DETECTION. PROCEDURE WAS COMPLETED AS NORMAL. WHEN MAT MALFUNCTION WAS IDENTIFIED, SPONGE COUNT WAS CONFIRMED WITH WAND AND PHYSICAL COUNT. THERE WAS NO PATIENT INJURY.