WM-NP2 WORKSTATION SET 5 (EU)
Received Apr 28, 2021 · Event occurred Apr 13, 2021
Report 9611174-2021-00014 · MDR key 11737160
Device
Generic name
Wm-*p2 Series Of Endoscopy Workstations
Manufacturer
Keymed (medical And Industrial Equipment) Ltd.Model number
K10021613Product problems
- Break
Patient
Not reported
- Insufficient Information
- No Clinical Signs, Symptoms or Conditions
- Insufficient Information
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
OLYMPUS KEYMED INVESTIGATION HAS BEEN COMPLETED. THE PHOTOS PROVIDED OF THE FAILURE SHOWED THAT THE CASTORS HAD RECEIVED MULTIPLE IMPACTS AND THAT THE LOCKING NUT AND HAD NOT BEEN SUFFICIENTLY TIGHTENED. THIS INDICATED THAT THE 6 MONTHLY CHECKS ON THE CASTORS HAD NOT BEEN CARRIED OUT AS PER THE INSTRUCTIONS FOR USE. FAILURE CAUSED BY USER AND LACK OF REGULAR MAINTENANCE.
Additional Manufacturer Narrative
OLYMPUS KEYMED INVESTIGATION IN PROGRESS. HAVE REQUESTED THAT THE FAILED CASTORS ARE RETURNED FOR FURTHER INVESTIGATION.
Description of Event or Problem
DURING A PROCEDURE THE CASTORS ON THE WM-NP2 WORKSTATION BROKE OFF. THERE WAS NO REPORT OF INJURY TO PATIENT OR USER. A SUPPORT WAS PLACED UNDER THE BASE AND THE PROCEDURE WAS COMPLETED.