CHROMOPHARE OR LIGHT CEILING-MOUNTED

Received Apr 16, 2021 · Event occurred Mar 2, 2021

Report 0008010153-2021-00001 · MDR key 11682269

Device

Generic name

Light, Surgical, Ceiling Mounted

Model number

CH00000001

Catalog number

CH00000001

Product problems

  • Mechanical Problem
  • Mechanical Problem

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

[SUPPLEMENTAL 001] FIELD D4 WAS UPDATED TO CORRECT SERIAL NUMBER. FIELD G6 WAS UPDATED TO 'FOLLOW-UP REPORT # 001.' FIELD H2 WAS UPDATED TO 'CORRECTION.'

Description of Event or Problem

IT WAS REPORTED THE OR 2 LIGHT WON'T CONTROL FROM THE WALL. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

Additional Manufacturer Narrative

IT WAS REPORTED THAT THERE WAS A LOSS OF CONTROL OF THE LIGHT HEAD. WHEN THE STRYKER FIELD SERVICE TECHNICIAN WENT ONSITE HE FOUND THAT THE SPANNER NUT AND C-LIP WERE MISSING. THERE WAS NO PATIENT INVOLVEMENT OR INJURY REPORTED; THE SEVERITY IS S0 PER (B)(4) . IT IS UNKNOWN HOW THE C-CLIP AND SPANNER NUT ARE MISSING; POTENTIAL REASONS INCLUDE INCORRECT ASSEMBLY DURING MANUFACTURING AND UNAUTHORIZED FIELD MODIFICATION. (B)(4) IS OPEN TO ADDRESS THIS ISSUE.

Description of Event or Problem

IT WAS REPORTED THE OR 2 LIGHT WON'T CONTROL FROM THE WALL. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.