Narrative
Additional Manufacturer Narrative
H6: INVESTIGATION SUMMARY THIS COMPLAINT IS FOR THE MIS-IDENTIFICATION OF KLEBSIELLA VARIICOLA AS KLEBSIELLA PNEUMONIAE. THERE ARE NO CLAIMS FOR THIS ORGANISM AS IT IS NOT IN THE PHOENIX ID DATABASE. THIS COMPLAINT IS NOT CONFIRMED. A REVIEW OF COMPLAINTS REVEALED NO ADDITIONAL COMPLAINTS ON THIS CATCH. A REVIEW OF QUALITY NOTIFICATIONS REVEALED THERE WERE NO QUALITY NOTIFICATIONS NOTED ON THIS BATCH.
Description of Event or Problem
IT WAS REPORTED WHILE USING BD BBL¿ CRYSTAL¿ ENTERIC/NONFERMENTER ID SYSTEM KLEBSIELLA VARIICOLA WAS MISIDENTIFIED AS K PNEUMONIAE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS UNEXPECTED ID OF KLEBSIELLA VARIICOLA AS K PNEUMONIAE NO FURTHER FOLLOW UP NEEDED, CUSTOMER WAS ADVISED THAT K VARIICOLA IS NOT IN DATABASE TAXA LIST FOR IDENTIFICATION.
Additional Manufacturer Narrative
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.
Description of Event or Problem
IT WAS REPORTED WHILE USING BD BBL¿ CRYSTAL¿ ENTERIC/NONFERMENTER ID SYSTEM KLEBSIELLA VARIICOLA WAS MISIDENTIFIED AS K PNEUMONIAE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS UNEXPECTED ID OF KLEBSIELLA VARIICOLA AS K PNEUMONIAE NO FURTHER FOLLOW UP NEEDED, CUSTOMER WAS ADVISED THAT K VARIICOLA IS NOT IN DATABASE TAXA LIST FOR IDENTIFICATION.