MalfunctionMRZ
MEDFUSION
Received Apr 8, 2021
Report 3012307300-2021-02973 · MDR key 11636926
Product problems
- Device Alarm System
- Power Problem
- Device Alarm System
- Power Problem
Patient
Not reported
- Insufficient Information
- Insufficient Information
Narrative
Additional Manufacturer Narrative
INVESTIGATION COMPLETED ON A SMITHS MEDICAL SYRINGE INFUSION PUMPS/MEDFUSION 3500 PUMPS . THE COMPLAINT OF WILL NOT BOOT UP COMPLETELY AND LIGHTS UP WHEN ALARMING WAS CONFIRMED. EVENT LOG REVEALED ERROR ALARM. WHEN POWERING UP DEVICE IT REVEALED A BLACK GREEN SCREEN AND ALARMED. THIS WAS ISOLATED TO THE MAIN BOARD WHICH WAS REPLACED.
Description of Event or Problem
INFORMATION RECEIVED A SMITHS MEDICAL SYRINGE INFUSION PUMPS/MEDFUSION 3500 PUMPS WOULD NOT BOOT UP COMPLETELY AND LIGHTS UP THEN STARTS TO ALARM. NO PATIENT INJURY REPORTED