MalfunctionPCL
CAPTIA MEASLES IGM ELISA KIT
Received Mar 29, 2021 · Event occurred Feb 10, 2021
Report 1318654-2021-00001 · MDR key 11584644
Device
Generic name
Measles Igm Elisa
Manufacturer
Trinity Biotech UsaModel number
2326060Catalog number
2326060
Lot number
060
Product problems
- Unable to Obtain Readings
- Unable to Obtain Readings
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
LABORATORY USER INFORMED MANUFACTURER THAT CAPTIA MEASLES IGM KIT WAS REPORTING INVALID ASSAYS ON TWO KITS FROM LOT 060 DUE TO POSITIVE CONTROL VALUES BEING ABOVE THE ASSIGNED ISR RANGE. NO INJURY WAS REPORTED.
Remedial action
- Replace
- Notification