MTS DILUENT 2 PLUS

Received Sep 4, 2008 · Event occurred Aug 6, 2008

Report 1056600-2008-00278 · MDR key 1156266

Device

Generic name

Red Blood Cell Diluent

Model number

NA

Catalog number

MTS9325

Lot number

012-189

Product problems

  • False Positive Result

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THAT THEY OBTAINED FALSE POSITIVE RESULTS WITH THE MTS A/B/D MONOCLONAL, AND REVERSE GROUPING CARD LOT # 101007037-27, WITH 2 PATIENT SAMPLES. THE CUSTOMER INDICATED THE PATIENTS WERE TYPED AS GROUP A POSITIVE, WITH THE ANTI-B AND CONTROL MICROTUBES APPEARING AS WEAKLY POSITIVE REACTIONS. THE CUSTOMER SUSPECTED THE MTS DILUENT 2 PLUS LOT # 012-189 USED IN PREPARATION OF THE CELL SUSPENSIONS WAS CONTAMINATED. NO ERRONEOUS RESULTS WERE REPORTED.

Additional Manufacturer Narrative

IT IS UNK HOW THE DILUENT BECAME CONTAMINATED. THE CUSTOMER INDICATED THEY OBTAINED ACCEPTABLE RESULTS WHEN THE SAMPLES WERE REPEATED USING THE SAME LOT OF GEL CARDS, AND A NEWLY OPENED DILUENT BOTTLE FROM THE SAME LOT. THE CUSTOMER ALSO INDICATED THAT THE DISPENSER, THAT WAS PREVIOUSLY INVOLVED IN TESTING, WAS NOT USED DURING REPEAT TESTING. THEREFORE, CONTAMINATED DISPENSER MAY HAVE CONTRIBUTED TO THIS INCIDENT. REVIEW OF MANUFACTURING DOCUMENTS INDICATES THAT THIS PRODUCT PASSED ALL RELEASE CRITERIA. THE CUSTOMER DID NOT REPORT OF PARTICULATES OR ANY ABNORMAL APPEARANCE IN THE DILUENT BOTTLE PRIOR TO USE. A COMPLAINT REVIEW INDICATED NO OTHER COMPLAINTS HAVE BEEN LOGGED AGAINST THIS LOT OF DILUENT. INCIDENT IS ISOLATED. (B) (4).