MalfunctionKOC

CLIC BLOOD CHAMBER

Received Mar 24, 2021 · Event occurred Mar 3, 2021

Report 8030665-2021-00404 · MDR key 11559433

Device

Generic name

Accessories, Blood Circuit, Hemodialysis

Model number

CL10041021

Catalog number

CL10041021

Product problems

  • Fluid/Blood Leak
  • Fluid/Blood Leak

Patient

72 YR

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PLANT INVESTIGATION: THE SAMPLE WAS NOT RETURNED TO THE MANUFACTURER AND THE LOT NUMBER WAS NOT PROVIDED. A MANUFACTURING REVIEW WAS PERFORMED ON THE PRODUCTS SHIPPED TO THE FACILITY FOR THE THREE (3) MONTH TIME FRAME WHICH IMMEDIATELY PRECEDED THE EVENT OCCURRENCE DATE. THIS REVIEW INCLUDED THE LOT NUMBERS FOR ALL CLIC BLOOD CHAMBERS SHIPPED TO THIS ACCOUNT WITHIN THE SELECTED TIME FRAME. THE ENTIRE SET OF LOTS HAVE BEEN SOLD AND DISTRIBUTED. THERE WERE NO NON-CONFORMANCES OR ABNORMALITIES IDENTIFIED DURING THE MANUFACTURING PROCESS WHICH COULD BE ASSOCIATED WITH THE REPORTED EVENT. AN INVESTIGATION OF THE DEVICE HISTORY RECORDS (DHR) WAS CONDUCTED AND CONFIRMED THAT THE RESULTS OF THE IN-PROGRESS AND FINAL QUALITY CONTROL (QC) TESTING MET ALL REQUIREMENTS. THE PRODUCT LOTS INVOLVED MET ALL SPECIFICATIONS FOR RELEASE. A REVIEW OF THE DHR DID NOT REVEAL A PROBABLE CAUSE FOR THE CUSTOMER COMPLAINT. AS A PHYSICAL EVALUATION COULD NOT BE PERFORMED, A DEFINITIVE CONCLUSION REGARDING THE REPORTED INCIDENT COULD NOT BE REACHED AND A CAUSE COULD NOT BE CONFIRMED.

Description of Event or Problem

A USER FACILITY CLINIC MANAGER (CM) REPORTED THAT A CLIC BLOOD CHAMBER BLOOD LEAK OCCURRED DURING A PATIENT¿S HEMODIALYSIS (HD) TREATMENT. THE CM CONFIRMED BLOOD WAS OBSERVED LEAKING EXTERNALLY FROM THE CLIC BLOOD CHAMBER WITHIN THE FIRST THIRTY MINUTES OF TREATMENT. THE CM WAS UNSURE WHICH END OF CLIC BLOOD CHAMBER THE LEAK WAS COMING FROM. HOWEVER, THE CM WAS ADAMANT THAT THE LEAK CAME FROM THIS DEVICE AND NOT FROM THE DIALYZER OR THE BLOODLINE. A CRIT-LINE REPRESENTATIVE HAPPENED TO BE ONSITE WHEN THE EVENT OCCURRED, AND THEY PROVIDED LIVE TROUBLESHOOTING SUPPORT. THE REPRESENTATIVE REALIZED THAT THE CLIC DEVICE WAS NOT FULLY ENGAGED. THEY TIGHTENED THE DEVICE AND THE LEAKING STOPPED. THE PATIENT WAS DIALYZING ON A B. BRAUN HD MACHINE, WITH A B. BRAUN STREAMLINE BLOODLINE AND A FRESENIUS OPTIFLUX DIALYZER. THERE WAS NO EVIDENCE OF ANY DAMAGE OR IRREGULARITIES FOUND ON THE CLIC DEVICE PRIOR TO OR AFTER USE. ADDITIONALLY, IT WAS UNKNOWN IF THERE WERE ANY MACHINE ALARMS DURING THE TREATMENT. THE CM STATED THEY ARE IN THE PROCESS OF RE-EDUCATING THE STAFF ON FULLY ENGAGING THE CLIC BLOOD CHAMBERS DURING SETUP, TO ENSURE THEY ARE SECURE. THE CM WAS UNSURE IF THE LEAKING WAS DUE TO OPERATOR ERROR OR IF THE CONNECTION WAS LOOSENED AS THE TUBING WARMED UP. AFTER THE CLIC DEVICE WAS TIGHTENED, THE PATIENT¿S TREATMENT WAS THEN CONTINUED AND COMPLETED WITHOUT ANY FURTHER ISSUES. THE CM STATED THE ESTIMATED BLOOD LOSS (EBL) WAS LESS THAN 5 ML. THE CM ALSO CONFIRMED THERE WAS NO PATIENT INJURY, NO ADVERSE EFFECTS WERE EXPERIENCED, AND NO MEDICAL INTERVENTION WAS REQUIRED AS A RESULT OF THE REPORTED EVENT. THE COMPLAINT DEVICE WAS DISCARDED AND THE LOT NUMBER IN USE COULD NOT BE DETERMINED.