MalfunctionKPJ
COMPRESSOR
Received Aug 25, 1997 · Event occurred Aug 3, 1997
Report MW1011935 · MDR key 115040
Device
Product problems
- Partial Blockage
- Malfunction
Patient
49 YR
- Unknown (for use when the patient's condition is not known)
Narrative
Description of Event or Problem
COMPRESSOR ON HBO CHAMBER DID NOT WORK NECESSITATING USE OF BACK-UP. PT WAS ABLE TO CONTINUE TREATMENT IN A SUCCESSFUL EFFORT TO SAVE AMPUTATED FINGERS. FAILURE OF COMPRESSOR HAS OCCURRED SEVERAL TIMES OVER LAST FEW MONTHS. FAILURE SEEMS TO BE RELATED TO SLUDGE IN THE OIL LINE POSSIBLY RELATED TO THE CHILLED WATER SYSTEM.