inforMED
OtherFII

PHOENIX

Received Apr 29, 2008 · Event occurred Mar 26, 2008

Report 2087532-2008-00044 · MDR key 1144032

Device

Generic name

Hemodialysis Equipment

Manufacturer

Gambro Dasco

Model number

PHOENIX

Catalog number

6023006700

Lot number

NA

Product problems

  • Device Alarm System
  • Dislodged
  • Fluid/Blood Leak
  • Normal

Patient

UNKNOWN

  • No Consequences Or Impact To Patient
  • Therapy/non-surgical treatment, aborted/stopped
  • Blood Loss

Narrative

Description of Event or Problem

GAMBRO RECEIVED A COMPLAINT ON MARCH 28TH REPORTED THAT A PT'S VENOUS NEEDLE BECAME DISLODGED APPROX 1.5 HOURS INTO A ROUTINE DIALYSIS TREATMENT ON A PHOENIX MACHINE, SOFTWARE VERSION 3.34. ACCORDING TO THE FACILITY'S TECHNICAL DIR, THE PT FELT WETNESS ON HIS SHIRT AND ALERTED THE NURSING STAFF. NO ALARMS WERE GENERATED ON THE PHOENIX MACHINE. A MINOR BLOOD LOSS WAS REPORTED AND THE TREATMENT WAS DISCONTINUED. NO MEDICAL INTERVENTION OR INJURY REPORTED. THE PT WAS DISCHARGED HOME AND RETURNED FOR HIS NEXT SCHEDULED HEMODIALYSIS TREATMENT IN 2008. THE MACHINE WAS INSPECTED BY A GAMBRO TECH AND RESULTED TO BE WORKING AS PER SPECS. THE TECH AUTHORIZED THE MACHINE TO BE PLACED BACK IN SVC.