inforMED
MalfunctionNQD

CRPHS CARDIAC C-REACTIVE PROTEIN (LATEX) HIGH SENSITIVE

Received Feb 26, 2021 · Event occurred Dec 9, 2020

Report 1823260-2021-00595 · MDR key 11387183

Device

Generic name

Cardiac C-reactive Protein, Antigen, Antiserum, And Control

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

04628918190

Lot number

470708

Product problems

  • High Test Results
  • High Test Results

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE QC RECOVERY WAS WITHIN SPECIFICATION. SAMPLES WERE REQUESTED FOR FURTHER INVESTIGATION BUT WERE NOT AVAILABLE. INVESTIGATIONS USING NATIVE SAMPLES WITH CRP CONCENTRATION ABOVE 100 MG/L COULD NOT CONFIRM THE ISSUE AND ALL RESULTS WERE WITHIN SPECIFICATIONS. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED. THIS EVENT OCCURRED IN (B)(6). UNIQUE IDENTIFIER (UDI) # (B)(4).

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS BETWEEN THE CRPL4 C-REACTIVE PROTEIN GEN.4 AND CRPHS CARDIAC C-REACTIVE PROTEIN HIGH SENSITIVE ASSAYS. FOR ONE PATIENT, THE CRP4 RESULT FROM COBAS 8000 COBAS C 701 MODULE SERIAL NUMBER (B)(4)WAS 72.3 MG/L. THE CRPHS RESULT WITH A 1:15 DILUTION FROM A COBAS 8000 C 502 MODULE WAS 136 MG/L. BOTH VALUES WERE REPORTED TO THE MEDICAL PERSONNEL. IT WAS REQUESTED BUT WAS NOT KNOWN WHICH RESULT WAS BELIEVED TO BE CORRECT.