STRAP RPLT ACTIVE HEEL TRACTION BOOT
Received Feb 3, 2021 · Event occurred Jan 13, 2021
Report 3003604053-2021-00029 · MDR key 11273900
Device
Generic name
Apparatus, Traction, Non-powered
Manufacturer
Smith & Nephew, Inc.Model number
72203275Catalog number
72203275
Lot number
UNKNOWN
Product problems
- Unintended Movement
- Unintended Movement
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INTERNAL COMPLAINT REFERENCE CASE: (B)(4).
Description of Event or Problem
IT WAS REPORTED THAT ACTIVE HEEL TRACTION BOOT REPLACEMENT KIT WAS LOOSENING AND IT WAS NOT MAINTAINING TENSION ON THE BOOT CAUSING THE PATIENT´S FOOT TO SLIDE, ULTIMATELY IMPACTING DISTRACTION. THIS WAS NOTICED BEFORE THE PROCEDURE; THEREFORE, NO PATIENT WAS INVOLVED. A BACKUP DEVICE WAS AVAILABLE AND NO DELAY WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Additional Manufacturer Narrative
H10: H3, H6: THE REPORTED DEVICE, INTENDED FOR USE IN TREATMENT, WAS NOT RETURNED TO THE DESIGNATED COMPLAINT UNIT FOR INDEPENDENT EVALUATION, THUS VISUAL INSPECTION AND FUNCTIONAL TESTING COULD NOT BE PERFORMED. A COMPLAINT HISTORY REVIEW CONCLUDED THIS WAS A REPEAT ISSUE. A RELATIONSHIP, IF ANY, BETWEEN THE SUBJECT DEVICE AND THE REPORTED EVENT COULD NOT BE DETERMINED. PLEASE REFER TO THE INSTRUCTIONS FOR USE FOR RECOMMENDATIONS ON PROPER USE OF THE DEVICE AND POTENTIAL TROUBLESHOOTING METHODS TO PREVENT FUTURE REOCCURRENCE OF THE REPORTED EVENT. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME. IF THE PRODUCT ASSOCIATED WITH THIS EVENT IS RETURNED AT A FUTURE DATE, THIS EVALUATION WILL BE REOPENED FOR INVESTIGATION.