WELLCOGEN BACTERIAL ANTIGEN KIT

Received Feb 3, 1994 · Event occurred Jan 10, 1994

Report MW1000631 · MDR key 11260

Device

Catalog number

ZL26

Lot number

3312J

Product problems

  • False Negative Result

Patient

3 YR

Narrative

Description of Event or Problem

A 3-YR-OLD CHILD WAS ADMITTED TO THE ER ON 1/10/94 WITH NAUSEA, VOMITING AND COMPLAINT OF A HEADACHE. THE CHILD ALSO HAD BLUE SPLOTCHES OVER HER ENTIRE BODY. A SPINAL TAP WAS PERFORMED ALONG WITH OTHER LAB WORK. THE LAB PERFORMED A CULTURE, GRAM STAIN, LATEX MENINGITIS PANEL, GLUCOSE, PROTEIN AND CELL COUNT ON THE CSF. 1) THE GLUCOSE AND PROTEIN WERE NORMAL. 2) ON THE CELL COUNT THE APPEARANCE WAS CLEAR AND COLORLESS, 5 WBC AND 6 RBC. 3) THE GRAM STAIN SHOWED NO PMNS AND NO BACTERIA SEEN. 4) THIS LATEX MENINGITIS PANEL SHOWED ALL 5 ANTIGENS WERE NEGATIVE. THEY INCLUDE STREPTOCOCCUS GROUP B, H. INFLUENZAE TYPE B, S. PNEUMONIAE, N. MENINGITIDIS A, B, C, Y, W135 AND E. COLI K1. 5) COAGULATION STUDIES WERE ALSO PERFORMED WITH ELEVATED PT, PTT AND FDP. 6) A LIVER PROFILE WAS DONE WITH ELEVATED ALK, PHOS, LD AND SGOT. 7) THE CBC HAD A WBC OF 2.8, A PLT COUNT OF 26 AND A REVERSED DIFF. THE PT WAS THEN TRANSFERED TO A PICU UNIT AT ANOTHER HOSP WHERE SHE DIED. THEY PREFORMED A LATEX MENINGITIS PANEL ON SERUM AT THE OTHER HOSP WHERE THEY GOT A POSITIVE N. MENINGITIDIS A, C, Y, W135. IT MUST BE NOTED THAT THE KIT USED WAS ALSO THIS CO'S BACTERIAL ANTIGEN KIT. THE FOLLOWING DAY RPTR WAS TOLD THAT THE CHILD HAD DIED OF NEISSERIA MENINGITIDIS. RPTR WAS CONCERNED BECAUSE THE GRAM STAIN SHOWED NO WBCS AND NO BACTERIA, ALSO THE CULTURES WERE NO GROWTH AT 14 HRS (BOTH BLOOD AND CSF CULTURES). ON INVESTIGATION RPTR DISCOVERED THAT THE HOSP IN WHICH THE CHILD WAS TRANSFERED PREFORMED A MENINGITIS ANTIGEN PANEL AND GOT A POSITIVE AGGLUTINATION TO THE N. MENINGIDITIS A, C, Y, W135. WITH THIS KNOWLEDGE, RPTR PULLED THE SPINAL FLUID RERUNING THE TECH'S SAMPLE, REPROCESSING A NEW SAMPLE, AND BECAUSE RPTR WAS IN A HURRY, RPTR RAN THE CSF STRAIGHT. THE RESULTS OF RPTR'S STUDY WERE: 1) THE SAMPLE PERFORMEDBY THE NIGHT TECH WAS AS SHE REPORTED: ALL NEGATIVE. 2) THE SAMPLE RPTR REPROCESSED WAS ALSO ALL NEGATIVE. 3) BUT THE SAMPLE THAT WAS NOT PROCESSED AND RUN STRAIGHT WAS POSITIVE FOR NEISSERIA MENINGITIDIS A, C, Y, W135. (THE CONTROL WAS NEGATIVE). 4) THE CSF CULTURE AT 22 HRS SHOWED GROWTH OF 3 COLONIES OF GRAM NEGATIVE DIPLOCOCCUS THAT WAS OXIDASE POSITIVE. 5) THE NEXT DAY SUGARS, WERE RUN WITH AN ID OF NEISSERIA MENINGITIDIS. 6) THE SPECIMAN WAS SENT TO THE STATE LAB FOR TYPING WITH A FINDING OF TYPE C. THE MFR WAS PHONED AS SOON AS RPTR NOTED THE PROBLEM.