UMBILICAL CORD CLAMP
Received Jan 19, 2021 · Event occurred Dec 16, 2020
Report 1057404-2020-00007 · MDR key 11195096
Device
Generic name
Device, Occlusion, Umbilical
Manufacturer
Deroyal Industries, Inc.Model number
10-1281Lot number
53179059
Product problems
- Mechanical Problem
- Detachment of Device or Device Component
- Mechanical Problem
- Detachment of Device or Device Component
Patient
1 DA
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
Narrative
Additional Manufacturer Narrative
ROOT CAUSE: BECAUSE THE DEVICE WAS NOT RETURNED FOR EVALUATION, THE ROOT CAUSE WAS UNABLE TO BE DETERMINED AT THIS TIME. BASED ON THE PRODUCT RISK ANALYSIS, THE FOLLOWING POTENTIAL ROOT CAUSES HAVE BEEN IDENTIFIED AND COULD NOT BE RULED OUT DURING THE INVESTIGATION: MANUFACTURING ERROR AND CLINICIAN ERROR. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS NOT BEEN TAKEN DUE TO THE ROOT CAUSE DETERMINATION. INVESTIGATION SUMMARY AN INTERNAL COMPLAINT ((B)(4)) WAS RECEIVED INDICATING THAT AN UMBILICAL CORD CLAMP (PART 10-1281, LOT 53179059) SLIPPED FROM THE UMBILICAL CORD DURING USE. THE PATIENT SUFFERED BLOOD LOSS AND REQUIRED A TRANSFUSION. A SAMPLE WAS NOT RETURNED FOR EVALUATION. THE DEVICE HISTORY RECORD WAS REVIEWED FOR DOCUMENTATION OF NONCONFORMITIES IN VISUAL INSPECTIONS. ALL INSPECTIONS WERE ACCEPTABLE. DURING THE MANUFACTURING OF THE REPORTED LOT, (B)(4) CLAMPS WERE FUNCTIONALLY TESTED TO ENSURE THEY CLOSE. NO ISSUES WERE REPORTED. ADDITIONALLY, THE REPORTED LOT WAS AT THE BEGINNING OF THE PRODUCTION RUN, AND THE PRESSURE TEST WAS PERFORMED WITH ACCEPTABLE RESULTS. AN INVENTORY EVALUATION WAS CONDUCTED TO ENSURE THERE WERE NO ABNORMALITIES IN THE LATCH AND TEETH AREA THAT COULD POTENTIALLY CONTRIBUTE TO THIS OCCURRENCE. THERE WERE NO ISSUES DETECTED DURING THE EVALUATION. THE RISK ANALYSIS FILE WAS REVIEWED FOR POTENTIAL CAUSES FOR THE CLAMP BECOMING DISLODGED FROM THE UMBILICAL CORD. THE FOLLOWING POTENTIAL CAUSES WERE IDENTIFIED: IMPROPER MATERIAL SELECTION, RAW MATERIAL DEFECTIVE OR OUT OF SPECIFICATION, DESIGN ERROR, MANUFACTURING ERROR, CLINICIAN ERROR OR MISAPPLICATION, SHIPPING/HANDLING ERROR, PACKAGING ERROR, AND STORAGE ERROR. THERE HAS BEEN NO CHANGE TO THE MATERIAL USED TO MANUFACTURE THE PRODUCT. THE CERTIFICATE OF ANALYSIS FOR THE RAW MATERIAL IS TO BE REVIEWED TO ENSURE THE CERTIFICATE STATES CONFORMANCE TO ESTABLISHED SPECIFICATIONS. THE RAW MATERIAL LOT NUMBERS USED TO MANUFACTURE THE REPORTED LOT WERE REVIEWED AND ALL ARE ACCEPTABLE PER THE RANGES ON THE CERTIFICATE. THE RAW MATERIAL LOTS USED IN THE REPORTED LOT WERE USED IN OTHER UMBILICAL CORD CLAMP PRODUCTION LOTS WITH NO EXTERNAL REPORTS AGAINST THOSE FINISHED LOT NUMBERS. THE CLAMPS WERE MANUFACTURED FROM 10/4/2020 TO 10/24/2020 AND SHIPPED FROM THE MANUFACTURING FACILITY BETWEEN 10/7/2020 TO 11/9/2020. STORAGE FOR THIS LOT WAS MINIMAL COMPARED TO NORMAL STORAGE TIMEFRAMES DUE TO INCREASED PRODUCT DEMAND AT THE TIME. THEREFORE, IT IS UNLIKELY STORAGE CONDITIONS CONTRIBUTED TO THE REPORTED EVENT. THERE WAS NO REPORT OF THE PACKAGING ARRIVING DAMAGED FOR THIS SHIPMENT; THUS, THIS INCIDENT IS UNRELATED TO SHIPPING OR HANDLING. THE DESIGN AND PACKAGING OF THE PRODUCT HAVE NOT BEEN CHANGED, SO THIS INCIDENT IS NOT THE RESULT OF A DESIGN OR PACKAGING ERROR. (B)(4). THE INVESTIGATION IS COMPLETE AT THIS TIME. IF NEW AND CRITICAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.
Description of Event or Problem
THE UMBILICAL CORD CLAMP DID NOT HOLD AND TENDED TO SLIP. THE PATIENT, A NEWBORN, SUFFERED SIGNIFICANT BLOOD LOSS AND RECEIVED A BLOOD TRANSFUSION.