FRESENIUS DIALYSIS DELIVERYS SYSTEM

Received Aug 13, 2008 · Event occurred Jul 14, 2008

Report 2937457-2008-00008 · MDR key 1118447

Device

Generic name

Hemodialysis Machine

Model number

2008K

Catalog number

NA

Lot number

NA

Product problems

  • Device Rinsing Issue

Patient

64 YR

  • Other (for use when an appropriate patient code cannot be identified)

Narrative

Description of Event or Problem

A DIALYSIS FACILITY REPORTED THAT A PATIENT C/O FEELING WARM AND FLUSHED A FEW MINUTES AFTER A HEMODIALYSIS TREATMENT WAS INITIATED. HER SYMPTOMS GOT WORSE AND SHE C/O SHORTNESS OF BREATH AS TREATMENT PROGRESSED. THEY TESTED THE DIALYSATE FOR RESIDUAL BLEACH AND IT TESTED NEGATIVE AT THE DRAIN BUT POSITIVE AT THE HANSEN CONNECTOR. TREATMENT WAS TERMINATED, AND SHE WAS TAKEN TO THE HOSPITAL FOR ER. THE FLUID (SALINE AND BLOOD) IN THE EXTRACORPOREAL CIRCUIT WAS BROWNISH IN COLOR.

Additional Manufacturer Narrative

DEVICE INVESTIGATION SHOWED THAT ONE OF THE CONCENTRATE WANDS WAS NOT PROPERLY INSERTED INTO THE RINSE PORT AFTER A CHEMICAL DISINFECTION. RESIDUAL CHEMICAL WAS PROBABLY NOT CLEARED FROM THE TUBING AND THE STAFF FAILED TO TEST THE MACHINE FOR RESIDUAL BLEACH PRIOR TO INITIATING TREATMENT.