AMSORB
Received Jan 11, 2021 · Event occurred Dec 22, 2020
Report 11146550 · MDR key 11146550
Device
Generic name
Absorbent, Carbon-dioxide
Manufacturer
Armstrong Medical Ltd.Model number
AMAB3801-8Lot number
230420F112
Product problems
- Leak/Splash
- Therapeutic or Diagnostic Output Failure
Patient
Not reported
- Insufficient Information
- Insufficient Information
Narrative
Description of Event or Problem
OVER THE LAST COUPLE MONTHS IT HAS BEEN REPORTED AT TWO OF OUR HOSPITALS THAT THE O E MEYER AMSORB CARBON DIOXIDE ABSORBERS HAVE BEEN LEAKING. THIS ISSUE IS IDENTIFIED WHEN THE ANESTHESIA GAS MACHINE DOES NOT PASS ITS LEAK TEST. THE CANISTER IS THEN REMOVED AND REPLACED WITH ANOTHER CANISTER, UNTIL THE LEAK TEST IS PASSED AND THE MACHINE IS WORKING APPROPRIATELY. THIS ISSUE HAS BEEN IDENTIFIED AT BOTH OUR HOSPITAL LOCATIONS AND SEEMS TO PERSIST ACROSS SEVERAL DIFFERENT LOT NUMBERS. THREE OF THE LOT NUMBERS THAT WERE IDENTIFIED TO HAVE THIS ISSUE INCLUDE 170620F51, 17062OF61, AND 230420F112. THE MANUFACTURER REPLACED SOME OF THE AFFECTED PRODUCT.