inforMED
MalfunctionKWL

REMEDY SPECTRUM GV HIP

Received Dec 30, 2020 · Event occurred Dec 8, 2020

Report 3010537287-2020-00009 · MDR key 11096300

Device

Generic name

Hip Spacer, Product Code: Kwl

Manufacturer

Osteoremedies, Llc

Catalog number

GVHDSM

Lot number

OR01041

Product problems

  • Break
  • Material Integrity Problem
  • Break
  • Material Integrity Problem

Patient

56 YR

  • Failure of Implant
  • Limited Mobility Of The Implanted Joint
  • Failure of Implant
  • Limited Mobility Of The Implanted Joint

Narrative

Additional Manufacturer Narrative

THE USE OF CEMENT ON THE THREADS CONTRIBUTED TO THE DIFFICULTY AFFIXING THE HEAD TO THE STEM. THE EXTRA FORCE USED TO THREAD THE HEAD CAUSED THE HEAD IMPLANT TO BREAK. NO PATIENT HARM WAS REPORTED ONLY A DELAY IN SURGERY OF GREATER THAN 60 MINUTES.

Description of Event or Problem

WHEN AFFIXING THE HEAD TO THE STEM PRIOR TO IMPLANTATION, THE SURGEON CHOSE TO USE BONE CEMENT ON THE THREADS. DUE TO THE BONE CEMENT SETTING UP THE SURGICAL TECHNICIAN HAD DIFFICULTY THREADING THE HEAD ONTO THE STEM. THE SURGEON ALSO ATTEMPTED TO THREAD THE HEAD DOWN AND DUE TO AMOUNT OF FORCE BEING APPLIED THE HEAD IMPLANT BROKE. THE SURGERY WAS DELAYED FOR OVER 60 MINUTES WHILE ANOTHER HEAD IMPLANT WAS OBTAINED. THERE WAS NO REPORTED ADVERSE EVENT REPORTED FOR THE PATIENT DUE TO THE SURGICAL DELAY.