inforMED
InjuryFII

FRESENIUS DIALYSIS DELIVERY SYSTEM

Received Aug 5, 2008 · Event occurred Jul 4, 2008

Report 2937457-2008-00007 · MDR key 1104503

Device

Generic name

Hemodialysis Machine

Model number

2008K

Catalog number

NA

Lot number

NA

Product problems

  • Device Alarm System

Patient

38 YR

  • Chest Pain
  • Breathing difficulties
  • Hospitalization required

Narrative

Description of Event or Problem

A DIALYSIS FACILITY REPORTED THAT TOWARDS THE END OF A HEMODIALYSIS TREATMENT, THE NURSE NOTICED THAT THE BLOODLINES WERE FILLED WITH AIR FROM THE ARTERIAL DRIP CHAMBER TO THE PT AND THERE WAS NO LEVEL DETECTOR ALARM. THE MACHINE THEN GAVE A VENOUS PRESSURE ALARM AS SOON AS THE PROBLEM WAS OBSERVED. THE PT C/O SHORTNESS OF BREATH AND CHEST PAIN. THE PT WAS PLACED IN TRENDELENBERG POSITION AND OXYGEN WAS ADMINISTERED. THE PT WAS TAKEN TO THE HOSPITAL. SHE HAS BEEN DISCHARGED SINCE AND HAS RESUMED OUTPATIENT HEMODIALYSIS TREATMENTS.

Additional Manufacturer Narrative

DEVICE WAS CHECKED BY THE FACILITY TECHNICIAN AND BY FRESENIUS. NO PROBLEM WAS FOUND IN BOTH INVESTIGATIONS. THE LEVEL DETECTOR WILL BE RETURNED TO THE MANUFACTURER FOR FURTHER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF INVESTIGATION REVEALS NEW INFO.