inforMED
MalfunctionIDY

TRIAGLE BIOMEDICAL SCIENCES

Received Dec 23, 1993 · Event occurred Dec 15, 1993

Report 10983 · MDR key 10983

Device

Generic name

Histology Floatation Work Station

Model number

H-FWS

Catalog number

NI

Lot number

NI

Patient

NA

Narrative

Description of Event or Problem

THE DEVICE WAS NOT BEING USED AND THE FLOTATION DISH HAD BEEN REMOVED FROM THE HEATING WELL. THE HEATING UNIT HAD BEEN ACCIDENTALLY TURNED ON AND THE PLASTIC FRAME OF THE DEVICE WAS STARTING TO MELT AND SMOKE WHILE IT WAS POSITIONED ON A LAB COUNTER. FORTUNATELY, A LAB TECH WORKING IN THE AREA SMELLED THE BURNING PLASTIC AND INVESTIGATED. THE DEVICE WAS UNPLUGGED AND TRANSPORTED TO OUR BIOMED DEPT FOR INSPECTION.