inforMED
MalfunctionIZZ

COMPAX40E

Received Aug 4, 2008 · Event occurred Jul 5, 2008

Report 2126677-2008-00055 · MDR key 1096911

Device

Model number

2225881

Catalog number

NA

Lot number

NA

Product problems

  • Locking mechanism failure

Patient

NA

  • No Patient Involvement

Narrative

Description of Event or Problem

IT WAS REPORTED THE TABLE LOCKS DID NOT ACTUATE, CAUSING THE TABLETOP TO MOVE IN TWO AXES WITHOUT RESISTANCE. THERE WAS NEITHER PATIENT INVOLVEMENT NOR INJURY REPORTED. THIS SITUATION COULD CONTRIBUTE TO AN INJURY IF A PATIENT OR OPERATOR WERE UNAWARE OF THIS CONDITION WHILE LOADING A PATIENT. THE ENSUING INSTABILITY COULD LEAD TO A FALL.

Additional Manufacturer Narrative

THE GE FIELD ENGINEER (FE) EVALUATED THE SYSTEM AND FOUND A BLOWN FUSE IN THE TABLE LOCK CIRCUIT. THE FE REPLACED THE FUSED, ADJUSTED THE FOOTSWITCH ASSEMBLY, AND VERIFIED THAT THE TABLE LOCKS WERE PERFORMING ACCORDING TO SPECIFICATIONS.