inforMED
MalfunctionCDD

ARCHITECT B12 REAGENT KIT

Received Dec 5, 2020 · Event occurred Nov 9, 2020

Report 3005094123-2020-00273 · MDR key 10951364

Device

Generic name

Radioassay, Vitamin B12

Manufacturer

A.i.d.d Longford

Model number

7K61-37

Catalog number

07K61-37

Lot number

12324UI00

Product problems

  • High Test Results
  • High Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem

THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT B12 RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE IS 211-911 PG/ML): PATIENT # INITIAL RESULT, ON (B)(6) 2020, WAS 240, REPEATS WERE 197 AND 160 PG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Additional Manufacturer Narrative

THE COMPLAINT INVESTIGATION FOR FALSELY ELEVATED ARCHITECT B12 RESULTS INCLUDED A SEARCH FOR SIMILAR COMPLAINTS, AND THE REVIEW OF COMPLAINT TEXT, TRENDING DATA, LABELING, DEVICE HISTORY RECORDS. TRENDING REVIEW DETERMINED NO RELATED TREND FOR THE PRODUCT. MANUFACTURING DOCUMENTATION FOR THE LIKELY CAUSE LOT WAS REVIEWED AND DID NOT IDENTIFY ANY ISSUES. DEVICE HISTORY RECORD REVIEW ON LOT NUMBER 12324UI00 DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS WITH THE LIKELY CAUSE LOT. RETURN TESTING WAS NOT COMPLETED, AS RETURNS WERE NOT AVAILABLE. LABELING WAS REVIEWED AND FOUND TO ADEQUATELY ADDRESS THE ISSUE UNDER REVIEW. THE OVERALL PERFORMANCE OF ARCHITECT B12 REAGENTS IN THE FIELD WAS REVIEWED USING DATA GATHERED VIA ABBOTTLINK FROM CUSTOMERS WORLDWIDE AND SUGGESTED THAT THE PERFORMANCE IS ACCEPTABLE. BASED ON THE INFORMATION PROVIDED AND ABBOTT DIAGNOSTICS¿ COMPLAINT INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF ARCHITECT B12, LOT NUMBER 12324UI00, WAS IDENTIFIED.