ROCHE 9180 ELECTROLYTE ANALYZER
Received Nov 30, 2020 · Event occurred Nov 5, 2020
Report 1823260-2020-03036 · MDR key 10921053
Device
Generic name
Electrolyte Analyzer
Manufacturer
Roche DiagnosticsModel number
ISE 9180Catalog number
03157334001
Lot number
NA
Product problems
- Output Problem
- Output Problem
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
THE POTASSIUM ELECTRODE LOT NUMBER WAS 21500348 WITH AN EXPIRATION DATE OF 31-JUL-2020. THE INVESTIGATION CONFIRMED THE CUSTOMER PERFORMED THE RECOMMENDED MAINTENANCE PROCEDURES ON THE SYSTEM. THE INVESTIGATION REQUESTED THE AFFECTED SNAPPAK, REFERENCE ELECTRODE, POTASSIUM ELECTRODE, AND NEEDLE UNIT FOR FURTHER INVESTIGATION. THE CUSTOMER WAS UNABLE TO SEND THE PARTS IN FOR AN INVESTIGATION AS THE PARTS REMAINED IN USE. THE CUSTOMER HAS NOT REPORTED ANY FURTHER ISSUES. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
Additional Manufacturer Narrative
THIS EVENT OCCURRED IN (B)(6). THE CUSTOMER'S QC WAS ACCEPTABLE. THE FIELD SERVICE ENGINEER CONFIRMED THE CUSTOMER'S ANALYZER WAS OK AND DIDN'T FIND ANY ABNORMALITIES. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THE INITIAL REPORTER RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED FOR POTASSIUM ON THE ROCHE 9180 ELECTROLYTE ANALYZER. THE PATIENT'S POTASSIUM RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER PERFORMED REPEAT TESTING WITH THE PATIENT'S SAMPLE FOR CONFIRMATION. AT 12:40 P.M., THE PATIENT'S POTASSIUM RESULT WAS 5.4 MMOL/L WITH A DATA FLAG. AT 12:42 P.M., THE PATIENT'S POTASSIUM RESULT WAS 5.6 MMOL/L WITH A DATA FLAG. AT 12:59 P.M., THE PATIENT'S POTASSIUM RESULT WAS 5.3 MMOL/L WITH A DATA FLAG. AT 13:01 P.M., THE PATIENT'S POTASSIUM RESULT WAS 4.5 MMOL/L. AT 13:02 P.M., THE PATIENT'S POTASSIUM RESULT WAS 5.3 MMOL/L WITH A DATA FLAG. THE CUSTOMER PERFORMED AN ADDITIONAL POTASSIUM MEASUREMENT, AND THE PATIENT'S POTASSIUM RESULT WAS 5.1 WITH A DATA FLAG. THE TIME OF MEASUREMENT WAS REQUESTED BUT NOT PROVIDED. THE POTASSIUM ELECTRODE LOT NUMBER WAS 21500348 WITH AN EXPIRATION DATE REQUESTED BUT NOT PROVIDED. THE POTASSIUM ELECTRODE HAD AN INSTALL BY DATE OF(B)(6) 2020, AND THE ELECTRODE WAS INSTALLED ON (B)(6) 2020.