CYMEDICA ORTHOPEDICS E-VIVE
Received Nov 30, 2020 · Event occurred Oct 29, 2020
Report 3011466827-2020-00004 · MDR key 10920525
Device
Generic name
Cymedica Orthopedics® E-vive¿
Manufacturer
Cymedica Orthopedics, Inc.Model number
CY-1000-310Catalog number
CY-1000-310
Lot number
0214885-0147
Product problems
- Use of Device Problem
- Improper or Incorrect Procedure or Method
- Use of Device Problem
- Improper or Incorrect Procedure or Method
Patient
55 YR
- Superficial (First Degree) Burn
- Superficial (First Degree) Burn
Narrative
Additional Manufacturer Narrative
THE OBSERVED TRAUMA IS INDICATIVE OF AN ACUTE LOCALIZED INCREASE IN CURRENT DENSITY AS A RESULT OF LOSS OF PAD CONDUCTIVITY ACROSS ITS SURFACE AREA. MANY POTENTIAL SOURCES EXIST OF ACCIDENTAL HIGH CURRENT DENSITIES. WRINKLES OR BREAKS IN ELECTRODE CONDUCTORS, NON-UNIFORM GEL LAYER OR DAMAGE, DRYING OUT, ELECTRODES PARTIALLY PEELING OFF THE SKIN (LOSS OF ADHESION), INCORRECT ELECTRODE PLACEMENT AND ELECTRODE DESIGN CAN CREATE PREFERENTIAL CURRENT THROUGH A REDUCED SURFACE AREA INTO THE PATIENT RESULTING UNSAFE CURRENT DENSITIES AND SUBSEQUENT TISSUE DAMAGE. THIS CAN OCCUR EVEN AT SAFE THERAPEUTIC IMPULSES DUE TO THE REDUCTION OF THE CONDUCTIVE SURFACE AREA RESULTING IN UNSAFE CURRENT DENSITIES. IN THE ABSENCE OF THE PHYSICAL COMPLAINT DEVICE TO INVESTIGATE, PHOTO AND TESTIMONY EVIDENCE WAS INVESTIGATED AS FOLLOWS: THE COMPLAINT DEVICE ELECTRODE PHOTOS WERE EXAMINED AS FOLLOWS: WRINKLES: NO SIGN OF DAMAGE. BREAKS: NO SIGN OF DAMAGE. GEL: SUBSTANTIAL PERIMETER DAMAGE INDICATIVE OF EXCESSIVE USE AND POOR MAINTENANCE DRYNESS: PHOTOS ARE COMMENSURATE WITH DRY ELECTRODES. ADHESION: PHOTOS ARE COMMENSURATE WITH COMPROMISED ADHESION DUE TO DRYNESS. INCORRECT ELECTRODE PLACEMENT: UNKNOWN, PHOTOS OF INJURIES SUGGEST CORRECT PLACEMENT ON LEG. THE COMPLAINT DEVICE WAS EXAMINED FOR: GARMENT INSPECTED, NO ABNORMALITIES. FAULT CONDITIONS, NO OVERVOLTAGE FAULTS REPORTED. ROOT CAUSES: PATIENT NEGLIGENCE OF REPLACING DRY ELECTRODES. AS THE ELECTRODES ARE USED, MOISTURE IS LOST DUE TO THERMAL HEATING. CONSEQUENTLY, INCREASING RESISTANCE, WHICH THEN REQUIRES ELEVATED INTENSITY LEVELS TO ACHIEVE STRONG CONTRACTIONS. THIS LEADS TO A GREATER ELECTRICAL POTENTIAL AND THUS A PROPENSITY OF HIGHER CURRENT DENSITIES. PATIENT NEGLIGENCE OF USING ELECTRODE GEL TO FACILITATE CURRENT DENSITY DISTRIBUTION AT HIGHER THERAPY LEVELS.
Description of Event or Problem
ON (B)(6) 2020 PATIENT CONTACTED CYMEDICA CUSTOMER SERVICE REGARDING BURNS HE RECEIVED FROM THE ELECTRODES. THE "BURN" MARKS ARE HEALED NOW; PLEASE TAKE NOTE THAT HE USED THE SAME SET OF PADS SINCE LATE AUGUST. THE PATIENT DID NOT USE GEL. CUSTOMER SERVICE SHIPPED HIM NEW ELECTRODES AND RE-EDUCATED HIM ON USE/CARE ETC.
Remedial action
- Patient Monitoring