inforMED
MalfunctionEIY

ROOT FILLER -L-, STERILE

Received Nov 29, 2020

Report 9611053-2020-00367 · MDR key 10916508

Device

Generic name

Instrument, Filling, Plastic, Dental

Manufacturer

Dentsply Vdw Gmbh

Model number

NA

Catalog number

V040393025030

Lot number

264710

Product problems

  • Break
  • Material Separation
  • Break

Patient

Not reported

  • Insufficient Information
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

THE INVOLVED PASTE FILLERS THAT HAVE BROKEN DURING USE HAVE NOT BEEN RETURNED AND CANNOT BE ANALYZED. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHRS REVIEW (BATCHES #1455636, #1478554 AND #1477871). THE UNUSED PASTE FILLERS 25MM N°2 RETURNED HAVE BEEN MEASURED AND HAVE BEEN FOUND IN COMPLIANCE WITH SPECIFICATIONS. FOR INFORMATION, NO MECHANICAL TEST IS APPLICABLE REGARDING THIS PRODUCT. ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. MULTIPLE UNSUCCESSFUL ATTEMPTS WERE MADE TO OBTAIN THE PATIENT OUTCOME.

Additional Manufacturer Narrative

THERE HAS BEEN A PREVIOUS REPORT WITH THE SAME PRODUCT WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH RESULTS ARE NOT AVAILABLE AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Description of Event or Problem

IN THIS EVENT IT WAS REPORTED THAT PASTE FILLERS OF A CERTAIN LOT ARE SEPARATING. NO INJURY REPORTED. THE OUTCOME OF THE EVENT IS UNKNOWN AS OF THIS MDR EVALUATION.