MalfunctionIOY
SUPPORT ARM 179
Received Nov 24, 2020
Report 8010042-2020-01063 · MDR key 10895986
Device
Generic name
Support, Arm
Manufacturer
Maquet Critical Care AbModel number
SUPPORT ARM 179Catalog number
6882001
Product problems
- Mechanical Problem
- Mechanical Problem
Patient
Not reported
- No Patient Involvement
- No Patient Involvement
Narrative
Additional Manufacturer Narrative
EXAMINATION OF THE RETURNED SUPPORT ARM FOUND THAT THE THREAD LOCKER FOR THE LOCKING/UNLOCKING SCREW HAD RELEASED AND THE JOINT COULD FALL APART. THE CAUSE HAS NOT BEEN DETERMINED BUT THE MOST PROBABLE CAUSE WAS FURTHER SCREWING/UNSCREWING OF THE HANDLE AFTER THE STOP OF LOCKING/UNLOCKING CAUSED THE DAMAGE.
Description of Event or Problem
IT WAS REPORTED THAT THE JOINT OF THE SUPPORT ARM WAS DEFECTIVE. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).