inforMED
MalfunctionFBO

INNER CERAMIC INSULATION TUBE

Received Nov 19, 2020 · Event occurred Oct 1, 2020

Report 9610617-2020-00128 · MDR key 10870740

Device

Generic name

Resectoscope Sheath, Inner Fixed Sheath

Model number

27040XB

Catalog number

27040XB

Lot number

XX

Product problems

  • Material Separation
  • Material Separation

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE CERAMIC TIP FRACTURED/BROKE FROM THE SHEATH. THE IMAGES SHOWED THAT A CERAMIC BEAK WAS NEW AND SHINNING AND IN DARK COLOR THAN THE FACTORY ORIGINAL, WHILE THE METAL SHEATH LOOKS AGED AND WITH DENTS. WE BELIEVE THE SHEATH WAS SERVICED BY A THIRD PARTY, AND THE BEAK WAS REPLACED WITH A DIFFERENT CERAMIC MATERIAL.

Description of Event or Problem

ALLEGEDLY, DURING A UROLOGY PROCEDURE, THE CERAMIC TIP OF THE SHEATH BROKE. THE PIECES WERE RETRIEVED, AND THE PROCEDURE WAS COMPLETED.