IRRIGATION KIT - CO2 LONG

Received Nov 18, 2020 · Event occurred Oct 16, 2020

Report 1528319-2020-00039 · MDR key 10864268

Device

Generic name

Irrigation Kit

Model number

00711559

Catalog number

00711559

Lot number

2006647

Product problems

  • Fluid/Blood Leak
  • Device Damaged by Another Device

Patient

Not reported

  • Insufficient Information
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

THROUGH FOLLOW UP WITH THE USER FACILITY, THE LEAK REPORTEDLY OCCURRED ON THE WATER TUBING ADJACENT TO A PINCH CLAMP. THE PINCH CLAMP WAS REPORTED TO HAVE BEEN MOVED ALONG THE TUBING WHILE IN A CLOSED POSITION. SUCH MOVEMENT WOULD PLACE ABNORMAL STRESS ON THE TUBING AND COULD CAUSE THE REPORTED EVENT TO OCCUR. STATEMENTS FROM THE INSTRUCTIONS FOR USE INCLUDE: "PRIOR TO CLINICAL USE, INSPECT AND FAMILIARIZE YOURSELF WITH THE DEVICE AND REVIEW THE DIAGRAM BELOW. IF THERE IS EVIDENCE OF DAMAGE, DO NOT USE THIS PRODUCT AND CONTACT YOUR LOCAL PRODUCT SPECIALIST. BOTH PINCH CLAMPS SHOULD BE IN THE OPEN POSITION WHEN REMOVED FROM THE PACKAGE." THE DEVICE SUBJECT OF THIS COMPLAINT WAS NOT RETAINED FOR EVALUATION. HOWEVER, THE USER FACILITY RETURNED ANOTHER DEVICE FROM THE SAME LOT FOR EVALUATION. TESTING OF THE RETURNED UNIT WAS UNABLE TO DUPLICATE THE REPORTED EVENT. THE DEVICE HISTORY RECORD WAS REVIEWED AND CONFIRMED THE DEVICE LOT WAS MANUFACTURED TO SPECIFICATION. THERE HAVE BEEN NO OTHER COMPLAINTS ASSOCIATED WITH THIS LOT. STERIS ENDOSCOPY PROVIDED IN-SERVICE TRAINING TO THE USER FACILITY ON THE PROPER USE AND SETUP OF THE IRRIGATION KIT. NO FURTHER ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM THEIR IRRIGATION KIT TUBING DAMAGING THE VIDEO PROCESSOR. THIS RESULTED IN A PROCEDURE DELAY. THERE WAS NO REPORTED HARM TO THE PATIENT OR USERS.