IRRIGATION KIT - CO2 LONG
Received Nov 18, 2020 · Event occurred Oct 16, 2020
Report 1528319-2020-00039 · MDR key 10864268
Device
Generic name
Irrigation Kit
Manufacturer
United States Endoscopy Group, Inc.Model number
00711559Catalog number
00711559
Lot number
2006647
Product problems
- Fluid/Blood Leak
- Device Damaged by Another Device
Patient
Not reported
- Insufficient Information
- Insufficient Information
Narrative
Additional Manufacturer Narrative
THROUGH FOLLOW UP WITH THE USER FACILITY, THE LEAK REPORTEDLY OCCURRED ON THE WATER TUBING ADJACENT TO A PINCH CLAMP. THE PINCH CLAMP WAS REPORTED TO HAVE BEEN MOVED ALONG THE TUBING WHILE IN A CLOSED POSITION. SUCH MOVEMENT WOULD PLACE ABNORMAL STRESS ON THE TUBING AND COULD CAUSE THE REPORTED EVENT TO OCCUR. STATEMENTS FROM THE INSTRUCTIONS FOR USE INCLUDE: "PRIOR TO CLINICAL USE, INSPECT AND FAMILIARIZE YOURSELF WITH THE DEVICE AND REVIEW THE DIAGRAM BELOW. IF THERE IS EVIDENCE OF DAMAGE, DO NOT USE THIS PRODUCT AND CONTACT YOUR LOCAL PRODUCT SPECIALIST. BOTH PINCH CLAMPS SHOULD BE IN THE OPEN POSITION WHEN REMOVED FROM THE PACKAGE." THE DEVICE SUBJECT OF THIS COMPLAINT WAS NOT RETAINED FOR EVALUATION. HOWEVER, THE USER FACILITY RETURNED ANOTHER DEVICE FROM THE SAME LOT FOR EVALUATION. TESTING OF THE RETURNED UNIT WAS UNABLE TO DUPLICATE THE REPORTED EVENT. THE DEVICE HISTORY RECORD WAS REVIEWED AND CONFIRMED THE DEVICE LOT WAS MANUFACTURED TO SPECIFICATION. THERE HAVE BEEN NO OTHER COMPLAINTS ASSOCIATED WITH THIS LOT. STERIS ENDOSCOPY PROVIDED IN-SERVICE TRAINING TO THE USER FACILITY ON THE PROPER USE AND SETUP OF THE IRRIGATION KIT. NO FURTHER ISSUES HAVE BEEN REPORTED.
Description of Event or Problem
THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM THEIR IRRIGATION KIT TUBING DAMAGING THE VIDEO PROCESSOR. THIS RESULTED IN A PROCEDURE DELAY. THERE WAS NO REPORTED HARM TO THE PATIENT OR USERS.