inforMED
InjuryFII

PHOENIX

Received Jul 11, 2008 · Event occurred May 15, 2008

Report 2087532-2008-00062 · MDR key 1080685

Device

Generic name

Hemodialysis Equipment

Manufacturer

Gambro Dasco

Model number

PHOENIX

Catalog number

6023006700

Lot number

NA

Product problems

  • Disconnection
  • Fluid/Blood Leak
  • Loose or Intermittent Connection

Patient

UNKNOWN

  • Burn(s)
  • Therapy/non-surgical treatment, additional
  • Burn, Thermal
  • Hospitalization required

Narrative

Description of Event or Problem

GAMBRO RECEIVED A REPORT FROM A CUSTOMER WHO COMPLAINED THAT THREE NURSES SUSTAINED BURNS WHEN CONCENTRATE TUBING CAME OFF OF THE BLUE MALE CONCENTRATE CONNECTOR ASSEMBLY INSERTED INTO THE LOWER FRONT PANEL ON A PHOENIX MACHINE. THE EVENTS OCCURRED ON THREE DIFFERENT MACHINES WHICH WERE ALL EQUIPPED WITH THE SOFTWARE VERSION 3.34. (FOR THE OTHER EVENTS REFER TO MDR 9616240-2008-00036 AND MFR 9616240-2008-00037). HOT WATER WAS SPRAYED ONTO THE RIGHT LOWER FOREARM OF THE FACILITY'S NURSE WHILE THE DEVICE WAS UNDERGOING HEAT DISINFECTION. AT THAT MOMENT, THE NURSE WAS WIPING DOWN THE MACHINE WITH A DILUTE BLEACH SOLUTION. THE NURSE WAS DIAGNOSED WITH FIRST- DEGREE BURNS AND WAS TREATED AT THE EMERGENCY DEPT AT THE HOSPITAL. THE MACHINE WAS INSPECTED BY THE CLINIC'S BIOMEDICAL TECHNICAN WHO FOUND THAT THE CONCENTRATE TUBING AND THE TUBING CLIP HAD COME LOOSE FROM THE MALE CONNECTOR ASSEMBLY. HE REPLACED THE TUBING AND THE CLIP BACK ONTO THE MALE CONNECTOR ASSEMBLY. THE MACHINE WAS RETURNED TO SERVICE WITHOUT ANY FURTHER TUBE DISCONNECTION EVENT.