PHOENIX
Received Jul 11, 2008 · Event occurred May 15, 2008
Report 2087532-2008-00062 · MDR key 1080685
Device
Generic name
Hemodialysis Equipment
Manufacturer
Gambro DascoModel number
PHOENIXCatalog number
6023006700
Lot number
NA
Product problems
- Disconnection
- Fluid/Blood Leak
- Loose or Intermittent Connection
Patient
UNKNOWN
- Burn(s)
- Therapy/non-surgical treatment, additional
- Burn, Thermal
- Hospitalization required
Narrative
Description of Event or Problem
GAMBRO RECEIVED A REPORT FROM A CUSTOMER WHO COMPLAINED THAT THREE NURSES SUSTAINED BURNS WHEN CONCENTRATE TUBING CAME OFF OF THE BLUE MALE CONCENTRATE CONNECTOR ASSEMBLY INSERTED INTO THE LOWER FRONT PANEL ON A PHOENIX MACHINE. THE EVENTS OCCURRED ON THREE DIFFERENT MACHINES WHICH WERE ALL EQUIPPED WITH THE SOFTWARE VERSION 3.34. (FOR THE OTHER EVENTS REFER TO MDR 9616240-2008-00036 AND MFR 9616240-2008-00037). HOT WATER WAS SPRAYED ONTO THE RIGHT LOWER FOREARM OF THE FACILITY'S NURSE WHILE THE DEVICE WAS UNDERGOING HEAT DISINFECTION. AT THAT MOMENT, THE NURSE WAS WIPING DOWN THE MACHINE WITH A DILUTE BLEACH SOLUTION. THE NURSE WAS DIAGNOSED WITH FIRST- DEGREE BURNS AND WAS TREATED AT THE EMERGENCY DEPT AT THE HOSPITAL. THE MACHINE WAS INSPECTED BY THE CLINIC'S BIOMEDICAL TECHNICAN WHO FOUND THAT THE CONCENTRATE TUBING AND THE TUBING CLIP HAD COME LOOSE FROM THE MALE CONNECTOR ASSEMBLY. HE REPLACED THE TUBING AND THE CLIP BACK ONTO THE MALE CONNECTOR ASSEMBLY. THE MACHINE WAS RETURNED TO SERVICE WITHOUT ANY FURTHER TUBE DISCONNECTION EVENT.