inforMED
MalfunctionFGY

VACUETTE PLASTIC CANNULA HOLDEX STERILE

Received Nov 2, 2020

Report 8020040-2018-00024 · MDR key 10771010

Device

Generic name

Tube Holder

Model number

450216

Catalog number

450216

Lot number

A18063GP

Product problems

  • Physical Resistance/Sticking

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

COMPLAINT (B)(4). RECEIVED 2 BAGS 450216/A18063GP FOR EVALUATION. WE HAVE NO FURTHER INVENTORY OF THE MATERIAL/BATCH. WE HAVE NO FURTHER COMPLAINTS ON THE MATERIAL/BATCH. WE FORWARDED THE COMPLAINT AND SAMPLES TO OUR AFFILIATED HEADQUARTERS IN (B)(4) FROM WHICH WE RECEIVE THIS PRODUCT. ACCORDING TO THEIR INVESTIGATION AND COMMENTS, A REVIEW OF PRODUCTION DOCUMENTATION OF THE CONCERNED MATERIAL/BATCH SHOWS NO IRREGULARITIES RELATED TO THE REPORTED ERROR. THE CUSTOMER RETURNED SAMPLES WERE VISUALLY CHECKED; NO DEVIATIONS WERE OBSERVED. SAMPLES WERE FUNCTIONALLY CHECKED; NO ABNORMALITIES WERE DETECTED. IN ADDITION, THE TENSILE STRENGTH TEST DID NOT REVEAL ANY DEVIATIONS. THE COMPLAINT COULD NOT BE CONFIRMED.

Description of Event or Problem