inforMED
MalfunctionGDZ

PILLING KNIFE HANDLE

Received Jul 3, 2008

Report 1044475-2008-00019 · MDR key 1069745

Device

Generic name

Knife Handle

Manufacturer

Teleflex Medical

Model number

NA

Catalog number

352950

Lot number

PP6 NN6

Product problems

  • Loose

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

INCIDENT REPORTED AS: BLADE WILL NOT SIT ONTO HANDLE PROPERLY. NO PATIENT AND/OR USER INJURY. NO INTERVENTION REPORTED.

Additional Manufacturer Narrative

INVESTIGATION REVEALED THAT A TOLERANCE STACK-UP CONDITION COULD PRODUCE INTERFERENCE FIT BETWEEN BLADE AND HANDLE. A CORRECTIVE ACTION WAS IMPLEMENTED AND REVISIONS TO THE DRAWINGS AND MANUFACTURED PARTS WAS COMPLETED.

Remedial action

  • Modification/Adjustment