inforMED
MalfunctionCGL

EPOC HOST

Received Oct 5, 2020 · Event occurred Sep 5, 2020

Report 3002637618-2020-00053 · MDR key 10631223

Device

Generic name

Epoc Host

Manufacturer

Epocal Inc.

Model number

HH-1R09-00-00

Catalog number

10736393

Product problems

  • Non Reproducible Results
  • Non Reproducible Results

Patient

14 YR

  • Cardiac Arrest
  • Cardiac Arrest

Narrative

Additional Manufacturer Narrative

SIEMENS HAS REQUESTED DATA FILES FOR INVESTIGATION BUT NONE HAVE BEEN PROVIDED. SINCE NO INSTRUMENT FILES COULD BE OBTAINED, LOT PERFORMANCE WILL BE ASSESSED. THE CUSTOMER IS CURRENTLY USING THIS EPOC DEVICE. THE CUSTOMER STATED THE PATIENT IS CURRENTLY STABLE. RESPONSE FROM CUSTOMER: "AFTER MUCH REVIEW AND TESTING, WE HAVE DETERMINED THE DISCREPANCY COULD ONLY HAVE BEEN CAUSED BY PREANALYTICAL ISSUES. IN RUNNING TEN SAMPLES, MY CORRELATION FOR THE POTASSIUM WAS QUITE TIGHT. WE HAVE CONCLUDED THE EPOC WAS NOT RESPONSIBLE FOR THE MI OF THAT PATIENT."

Description of Event or Problem

THE CUSTOMER REPORTED DISCREPANT POTASSIUM VALUES ON THE EPOC HOST WHEN COMPARED TO A NON-SIEMENS LAB INSTRUMENT.

Additional Manufacturer Narrative

SIEMENS HAS REQUESTED INSTRUMENT LOGS FROM THE CUSTOMER, HOWEVER THE CUSTOMER HAS NOT PROVIDED THE REQUESTED DATA. THE CAUSE OF THE K+ DISCREPANCIES COULD NOT BE DETERMINED FROM THE INFORMATION PROVIDED BY THE CUSTOMER. A REVIEW OF THE INHOUSE PERFORMANCE FOR THE CARD LOT USED, LOT 01-20118-10, DID NOT IDENTIFY ANY PRODUCT DEFICIENCIES. THE FAILURE RATE OF LOT 01-20118-10 IS NOT SHOWING AN INCREASED TREND IN THE FIELD, THEREFORE THERE IS NO FURTHER EVIDENCE THAT THE SYSTEM OR REAGENT CARDS ARE NOT PERFORMING AS INTENDED. IT SHOULD BE NOTED THAT SAMPLE HANDLING/DELAYS ARE KNOWN TO AFFECT SAMPLE RESULTS IF PROPER PRECAUTIONS ARE NOT CONSIDERED. THE CAUSE OF THIS EVENT IS UNKNOWN.