inforMED
MalfunctionLKN

SPECTRA IDL SET

Received Sep 29, 2020 · Event occurred Aug 27, 2020

Report MW5096954 · MDR key 10605937

Device

Generic name

Separator, Automated, Blood Cell And Plasma, Therapeutic

Manufacturer

Terumo Bct Inc.

Model number

SPECTRA IDL SET

Catalog number

12320

Lot number

20044035

Product problems

  • Contamination /Decontamination Problem

Patient

79 YR

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

PLASMA PUMP ALARMED DURING PROCEDURE A CELLULAR THERAPY COLLECTION PROCEDURE. PLASMA PUMP TUBING OVER TORQUED. FLUID NOTED WHEN DEVICE DOOR WAS OPENED. PROCEDURE WAS HALTED UNDER THE GUIDANCE OF THE MD AND JUNO DUE TO POSSIBLE CELLULAR THERAPY PRODUCT (CTP) CONTAMINATION. THE WHOLE BLOOD PROCESSED WAS 2.948 LITERS WITH A COLLECTED CTP VOLUME OF 30 ML. NO RINSEBACK WAS PERFORMED. THE OPTIA SPECTRA KIT LOT # 20044036. DEVICE # 6 (SERIAL#: (B)(4)). FDA SAFETY REPORT ID# (B)(4).