PEPGEN P-15
Received Jun 12, 2008 · Event occurred May 12, 2008
Report 3006146445-2008-00048 · MDR key 1060436
Device
Product problems
- Failure to Osseointegrate
Patient
44 YR
- Impaired Healing
Narrative
Additional Manufacturer Narrative
FAILURE OF BONE GRAFTS TO OSSEOINTEGRATE IS AN INHERENT RISK OF THE PROCEDURE, ALTHOUGH UNCOMMON. OTHER VARIABLES, BEYOND PRODUCT NOT PERFORMING TO SPECIFICATIONS OR BEING NON-STERILE, TO CONSIDER IN A DIFFERENTIAL DIAGNOSIS WOULD BE PT HEALTH AND SOCIAL HISTORY (PARTICULARLY SMOKING, WHICH IS A KNOWN RISK FACTOR FOR BOTH IMPLANT AND GRAFT FAILURE), SITE PREPARATION TRAUMA (HEAT OR PRESSURE) RESULTING IN NECROSIS OR SCARRING AND FIBROSIS, INSUFFICIENT PRIMARY STABILITY (SECONDARY TO BONE QUALITY AND/OR EXCESSIVE PREPARATION OF THE OSTEOTOMY AND QUANTITY OF MAXILLARY RIDGE BELOW THE MAXILLARY SINUS) OF THE IMPLANT RESULTING IN MICRO OR MACRO MOVEMENT AND SUBSEQUENT FIBROUS INTEGRATION, INDIVIDUAL GRAFTING TECHNIQUES, AND POST-OPERATIVE COMPLICATION (E.G. INFECTION), FROM THE INFO PROVIDED, IT IS NOT POSSIBLE TO DEFINITIVELY DETERMINE IF THE IMPLANT, GRAFT TECHNIQUE, PATIENT COMPLIANCE, POST-OPERATIVE COMPLICATION, ETC. WAS THE ETIOLOGY OF FAILURE HOWEVER, BECAUSE A SECOND SURGERY WAS REQUIRED TO REMOVE AND / OR REPLACE THE IMPLANT AND GRAFT, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY. THE IMPLANT LOSS WILL BE REPORTED VIA ASR, AS APPROPRIATE. THE LOT NUMBER WAS PROVIDED, THOUGH THE INVESTIGATION IS NOT COMPLETE AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
Description of Event or Problem
IT WAS REPORTED THAT PEPGEN P-15 WAS USED SIMULTANEOUSLY WITH IMPLANT PLACEMENT IN A PT WITH BONE QUALITY TYPE II AND FAIR ORAL HYGIENE; THE PT WAS ALSO REPORTED AS BEING A SMOKER. THE OSTEOTOMY WAS PREPARED VIA DRILLING AND BONE CONDENSING AND HEALING WAS TRANSGINGIVAL. THE IMPLANT DID NOT OSSEOINTEGRATE AND WAS LOST DURING HEALING, APPROXIMATELY THREE WEEKS POST-PLACEMENT. IT IS NOT CLEAR IF THE GRAFT WAS INTEGRATING WITH THE SURROUNDING VITAL BONE OR IF IT ALSO FAILED WITH THE IMPLANT.