inforMED
MalfunctionKSO

MTS 24 PLACE CENTRIFUGE

Received May 30, 2008 · Event occurred May 2, 2008

Report 1056600-2008-00184 · MDR key 1059223

Device

Generic name

Centrifuge

Model number

NA

Catalog number

MT515060

Lot number

NA

Product problems

  • Sticking

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THAT ALTHOUGH THE MTS 24 PLACE CENTRIFUGE WAS NOT SPINNING AND THE RPM WAS AT ZERO, THE TIMER WAS STILL COUNTING DOWN. INCORRECT CENTRIFUGATION TIME DURING TESTING, IF UNDETECTED, MAY LEAD ERRONEOUS TEST RESULTS.

Additional Manufacturer Narrative

A POSSIBLE ROOT CAUSE WAS DETERMINED. OCD CUSTOMER TECHNICAL SERVICES PERFORMED TROUBLESHOOTING WITH THE CUSTOMER VIA PHONE AND DETERMINED THAT THE CENTRIFUGE BELT WAS WORN. REPLACEMENT OF THE CENTRIFUGE BELT HAS RETURNED THE UNIT TO EXPECTED OPERATION. THIS CUSTOMER HAS NOT LOGGED ANY COMPLAINTS AGAINST THIS CENTRIFUGE SINCE THIS INCIDENT. (B) (4).