inforMED
MalfunctionIZZ

COMPAX 40E

Received Jun 6, 2008 · Event occurred May 8, 2008

Report 2126677-2008-00037 · MDR key 1057970

Device

Model number

2225881

Catalog number

NA

Lot number

NA

Product problems

  • Locking mechanism failure

Patient

NA

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

THE SITE REPORTED THAT THE COMPAX 40E TABLE LOCKS DID NOT ACTUATE, CAUSING THE TABLETOP TO MOVE IN TWO AXES WITHOUT RESISTANCE. THERE WAS NO INJURY REPORTED. THIS SITUATION COULD POTENTIALLY CONTRIBUTE TO A SERIOUS INJURY, IF A PT WERE TO BECOME UNSTEADY AND FALL.

Additional Manufacturer Narrative

THE GE FIELD ENGINEER (FE) EVALUATED THE SYSTEM, AND FOUND A DEFECTIVE TABLE POWER SUPPLY BOARD, WHICH PREVENTED POWER FROM BEING APPLIED TO THE TABLE LOCKS. THE FE REPLACED THE POWER SUPPLY BOARD, AND RETURNED THE PRODUCT TO SERVICE.