inforMED
MalfunctionFXM

AUTOCLAVABLE CAMERA HEAD

Received Sep 22, 2020

Report 8010047-2020-06726 · MDR key 10566719

Device

Generic name

Camera Head

Model number

OTV-S7H-1NA-10E

Product problems

  • Poor Quality Image
  • Poor Quality Image

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

THE LEGAL MANUFACTURER REVIEWED THE CONTENTS OF THIS COMPLAINT. THE LEGAL MANUFACTURER REPORTED THAT THE ROOT CAUSE HAS NOT BEEN IDENTIFIED. HOWEVER, THE LEGAL MANUFACTURER REPORTED THAT THE PROBABLE CAUSE FOR THE REPORTED EVENT WAS AS FOLLOWS: IT WAS SPECULATED THAT THE IMAGE NOISE OCCURRED BECAUSE THE CABLE WAS SUBJECTED TO UNEXPECTED STRESS DUE TO THE HANDLING OF THE USER AND THE CABLE UNIT FAILED. ALSO, THE BASE PLATE OF THE COUPLER WAS BENT DUE TO UNEXPECTED STRESS APPLIED TO THE HEAD, AND THE CENTER OF THE IMAGE WAS SHIFTED." WHEN CHECKING THE CONTENT OF THE INSTRUCTIONS FOR USE AT THE TIME OF SHIPMENT OF THE PRODUCTS RELATED TO IMPORTANT INFORMATION. PLEASE READ BEFORE USE, IT WAS CONFIRMED THAT THERE WERE THE FOLLOWING ENTRIES: NEVER EXCESSIVELY BEND, PULL, TWIST, COIL, SQUEEZE OR APPLY A CRUSHING FORCE TO THE CAMERA CABLE. THE CAMERA CABLE COULD BECOME DAMAGED. A DHR REVIEW WAS PERFORMED AND CONFIRMED THAT THERE WERE NO ABNORMAL RESULTS.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION. THE USER¿S COMPLAINT OF ¿BLACK IMAGE¿ WAS CONFIRMED. THE DEVICE WAS VISUALLY INSPECTED AND FOUND THAT THE IMAGE IS NOISY AND IT CUTS ON AND OFF DUE TO FAULTY CABLE UNIT. THE COUPLER BASE PLATE IS BENT CAUSING AN OFF-CENTERED IMAGE. THE LISTED PARTS WERE REPLACED. THE DEVICE IS FULLY FUNCTIONAL AND RETURNED TO THE USER FACILITY.

Description of Event or Problem

THE USER FACILITY REPORTED THAT THERE WAS AN ISSUE WITH THE IMAGE. THE IMAGE DOES NOT APPEAR. THERE IS A BLACK IMAGE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.