inforMED
MalfunctionHTT

ULTRAPOWER BUR, ROUND 3MM

Received Jun 3, 2008 · Event occurred Apr 28, 2008

Report 1017294-2008-00226 · MDR key 1056018

Device

Generic name

Bur

Manufacturer

Conmed Linvatec

Model number

NA

Catalog number

00702122600

Lot number

BBC49982

Product problems

  • Device Slipped
  • Manufacturing, Packaging or Shipping Problem

Patient

NA

  • No Patient Involvement

Narrative

Description of Event or Problem

OUR (B) (6) DISTRIBUTOR REPORTED THAT DURING RECEIVING AND INSPECTION OF THIS DEVICE, THEY FOUND THE BUR CAME LOOSE FROM THE CLAMSHELL INSIDE THE STERILE PACKAGING. THE DISTRIBUTOR DID NOT REPORT ANY COMPROMISE TO THE STERILITY OF THE PACKAGE.

Additional Manufacturer Narrative

INVESTIGATION FINDINGS: CONMED LINVATEC RECEIVED THIS BUR AND PACKAGING FOR EVAL. ON 05/29/2008, A VISUAL EXAMINATION OF THE PACKAGE CONFIRMED THE REPORT THAT THE BUR IS LOOSE FROM THE CLAMSHELL INSIDE THE STERILE PACKAGING. FURTHER INVESTIGATION FOUND SCRATCHES ON THE POUCH AND CONFIRMED A COMPROMISE TO THE STERILITY OF THE PACKAGING, MAKING THIS A POTENTIAL FOR INJURY AND A REPORTABLE EVENT. AN INVESTIGATION FOR THIS FAILURE MODE REMAINS IN PROCESS.