inforMED
MalfunctionCBC

SODASORB CANISTER

Received Sep 18, 2020

Report 3012307300-2020-09487 · MDR key 10552764

Device

Generic name

Bottle, Collection And Trap, Breathing System

Model number

008870

Catalog number

008870

Lot number

UNKNOWN

Product problems

  • Device Damaged Prior to Receipt by User
  • Device Damaged Prior to Receipt by User

Patient

Not reported

  • Chills
  • Chills

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RECEIVED ON 24-SEP-2020 AND FORWARDED TO THE SUPPLIER FOR INVESTIGATION. PER SUPPLIER RESPONSE IT HAS BEEN DETERMINED THAT THE DAMAGE OCCURS DURING SHIPMENTS OF SINGLE CARTONS OF PRE-PAKS® FROM THE DISTRIBUTORS TO END USERS. SMITHS MEDICAL HAS BEEN PROVIDED WITH SHIPPING PROCEDURES DETAILING PACKAGING THE SINGLE PACKS OF DISCS IN AN OVERPACK WITH SHIPPING POPCORN TO PREVENT DAMAGE TO THE DEVICE.

Description of Event or Problem

IT WAS REPORTED THAT TWO CASES OF SMITHS MEDICAL PREPAK CABISTERS CAME DAMAGED WHERE THE CANISTERS HAVE BEEN CRACKED AND THE SOLUTION IS LEAKING OUT. THERE WAS NO DAMAGE TO THE SHIPPING BOX. NO ADVERSE PATIENT EFFECTS WERE REPORTED.