CELL-DYN 22 WB CONTROL
Received May 27, 2008 · Event occurred May 27, 2008
Report MW5007045 · MDR key 1052883
Device
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
Not reported
- Unknown (for use when the patient's condition is not known)
Narrative
Description of Event or Problem
VALUE ASSIGNMENT ISSUES WITH AT LEAST 3 CONSECUTIVE LOTS OF CELL-DYN 22 QUALITY CONTROL MATERIAL PRODUCED BY ABBOTT LABORATORIES AND BEING USED FOR CELLDYN 3200 HEMATOLOGY ANALYZERS. CONTROLS RUN OUTSIDE MANUFACTURER RANGES FROM THE OUTSET DESPITE THE ANALYZER BEING IN CONTROL FROM THE PREVIOUS LOT AND BEING CALIBRATED ONLY A FEW MONTHS PREVIOUSLY. CALL CENTER STAFF ACKNOWLEDGE ISSUES, BUT NO FOLLOW-UP LETTERS ISSUED BY ABBOTT. CALIBRATION IS NOW RECOMMENDED BY CALL CENTER STAFF PER LOT OF QUALITY CONTROL MATERIAL DESPITE INSTRUMENT MANUAL STATING BIANNUAL IS ALL THAT IS NEEDED. NO EXPLANTATION GIVEN BY ABBOTT FOR POOR PRODUCT PERFORMANCE. POTENTIAL FOR ADVERSE EFFECT ON PATIENT DATA ESPECIALLY IF SEQUENTIAL RESULTS ARE MONITORED AS CALIBRATION MUST BE ADJUSTED APPROX EVERY 2 MONTHS OR WITH ISSUE OF NEW CONTROL LOT. CONCERN THAT NO OFFICIAL INFORMATION HAS BEEN PUBLISHED BY ABBOTT RECOGNIZING THE ISSUE AND PROVIDING STANDARDIZED RESOLUTION FOR CUSTOMERS, AND THE PATIENTS THEIR LABORATORY CUSTOMERS SERVE. DATES OF USE: 2008. DIAGNOSIS OR REASON FOR USE: CONTROL OF HEMATOLOGY LAB TESTING RESULTS.