inforMED
MalfunctionIZZ

COMPAX 40E

Received May 22, 2008 · Event occurred Apr 22, 2008

Report 2126677-2008-00035 · MDR key 1052509

Device

Model number

NI

Catalog number

NA

Lot number

NA

Product problems

  • System fails to activate
  • Locking mechanism failure

Patient

NA

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE COMPAX 40E TABLE TOP BRAKES WOULD NOT ACTUATE IN THE LATERAL DIRECTION, DUE TO A PROBLEM WITH THE MICRO SWITCH FOR THOSE BRAKES, CAUSING THE TABLETOP TO MOVE FREELY IN THE LATERAL DIRECTION. NO INJURY WAS REPORTED. THIS SITUATION COULD POTENTIALLY CONTRIBUTE TO A SERIOUS INJURY IF THE PT BECOMES UNSTEADY AND FALLS.

Additional Manufacturer Narrative

THE GE FIELD ENGINEER (FE) REPLACED THE FOOT PEDAL MICROSWITCH AND ADJUSTED THE TABLE LOCK BRAKES. AFTER VERIFYING THAT THE TABLE WAS OPERATIONAL, THE FE RETURNED THE SYS TO SVC.