inforMED
MalfunctionJKD

ELECSYS C-PEPTIDE

Received Sep 10, 2020 · Event occurred Aug 14, 2020

Report 1823260-2020-02232 · MDR key 10515500

Device

Generic name

C-peptides Of Proinsulin Test System

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

03184897190

Lot number

ASKU

Product problems

  • High Test Results
  • High Test Results

Patient

54 YR

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

ON (B)(6) 2020, THE PATIENT HAD A NEW SAMPLE COLLECTED. THE CUSTOMER PERFORMED ELECSYS INSULIN AND ELECSYS C-PEPTIDE TESTING ON A COBAS 8000 E 602 MODULE AND A SIEMENS IMMULITE 2000 ANALYZER. THE PATIENT'S RESULTS ON THE E 602 MODULE WERE INSULIN 57.45 UU/ML, C-PEPTIDE 1.44 NG/ML, MOLAR RATIO INSULIN/C-PEPTIDE 0.83, AND PEG RECOVERY 8%. THE PATIENT'S RESULTS ON THE SIEMENS IMMULITE 2000 ANALYZER WERE INSULIN 55.6 MCU/ML, C-PEPTIDE 1.05 NG/ML, AND MOLAR RATIO INSULIN/C-PEPTIDE 1.15. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE THE LABORATORY AND NO ADVERSE EVENTS OCCURRED.

Additional Manufacturer Narrative

INTERNAL METHOD COMPARISON DATA CONFIRM THE TENDENCY OF HIGHER RECOVERY OF THE ELECSYS C-PEPTIDE COMPARED TO SIEMENS IMMUNLITE C-PEPTIDE. IT NEEDS TO BE TAKEN INTO ACCOUNT THAT ASSAYS FROM DIFFERENT VENDORS CAN GENERATE DIFFERENT VALUES. THIS RELATES TO THE OVERALL SETUPS OF THE ASSAYS, THE ANTIBODIES USED AND, DIFFERENCES IN REFERENCE MATERIALS/METHODS AND THE STANDARDIZATION METHODOLOGY USED. ADDITIONALLY, SINGLE MEASUREMENTS CAN STATISTICALLY DEVIATE AND ARE NOT USEFUL TO DEMONSTRATE A GENERAL DIFFERENCE BETWEEN METHODS. FURTHER SAMPLE INVESTIGATIONS WERE NOT POSSIBLE BASED ON THE LACK OF REMAINING SAMPLE. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Additional Manufacturer Narrative

PHONE (B)(6). THIS EVENT OCCURRED IN (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS C-PEPTIDE RESULTS FOR ONE PATIENT TESTED ON A COBAS 8000 E 602 MODULE SERIAL NUMBER: (B)(4). THE CUSTOMER ALSO PERFORMED TESTING ON A SIEMENS IMMULITE 2000 ANALYZER TO COMPARE RESULTS AND TO DETECT ANY SAMPLE OR REAGENT INTERFERENCES. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE THE LABORATORY. REFER TO THE ATTACHMENT FOR ALL PATIENT DATA. THIS MEDWATCH IS FOR ELECSYS C-PEPTIDE. REFER TO MEDWATCH WITH PATIENT IDENTIFIER: (B)(6) FOR THE ELECSYS INSULIN ASSAY.