inforMED
OtherGWL

SANDMAN DIGITAL 20 (SD20)

Received May 16, 2008 · Event occurred Apr 16, 2008

Report 9680138-2008-00001_2 · MDR key 1051497

Device

Generic name

*

Manufacturer

Eb Neuro Spa

Model number

SD20 AMPL/HEADBOX

Catalog number

D.2118/D.2120

Lot number

*

Product problems

  • Unknown (for use when the device problem is not known)

Patient

*

  • Tingling
  • Burn, Thermal
  • Electric Shock
  • Treatment with medication(s)

Narrative

Description of Event or Problem

NELLCOR PURITAN BENNETT RECEIVED A REPORT FROM A SLEEP FACILITY THAT "A PT COMPLAINED OF SHOCKING FEELING THAT OCCURRED WHEN THE CUSTOMER CONNECTED THE PT HEADBOX TO THE HEADBOX CABLE. LATER, THAT EVENING THE PT COMPLAINED OF A TINGLING FEELING FROM THE ISO GROUND LEAD", BUT CONTINUED WITH THE STUDY THROUGHOUT THE NIGHT. IN THE MORNING, THE SLEEP FACILITY FURTHER REPORTED THAT THE PT COMPLAINED OF A BURN WHERE ONE OF THE LEADS HAD BEEN CONNECTED. THE PT WAS REFERRED TO AND TREATED BY A DERMATOLOGIST WHO PRESCRIBED ANTIBIOTIC CREAM.