inforMED
OtherJGY

ADVIA CHEMISTRY TRIGLYCERIDE METHOD

Received May 24, 2008 · Event occurred Feb 25, 2008

Report 2432235-2008-00060 · MDR key 1051308

Device

Generic name

Triglyceride Method

Model number

NA

Catalog number

B01-4133-01

Lot number

RGT. LOT # 095939

Product problems

  • Low Test Results

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

CUSTOMER REPORTED A DISCORDANT TRIGLYCERIDE RESULT WHEN USING REAGENT LOT # 095939. THE SAMPLE INITIALLY READ 483 MG/DL AND UPON RETEST WITH A DILUTED SAMPLE, THE CORRECTED VALUE WAS 2120 MG/DL. THE CUSTOMER BECAME AWARE OF THIS PROBLEM WHEN SHE NOTICED THAT THE REACTION CURVE FOR THE INITIAL RESULT WAS DISCREPANT. THERE WAS NO PATIENT INTERVENTION AS A RESULT OF THIS DISCORDANT TRIGLYCERIDE RESULT.

Additional Manufacturer Narrative

DISCUSSIONS WITH THE OEM REAGENT MANUFACTURER INDICATES THAT THE ADVIA CHEMISTRY TRIGLYCERIDE METHOD KIT LOT 095939 IS EXHIBITING LOWER THAN EXPECTED LINEARITY. THE ROOT CAUSE. INVESTIGATION IS UNDER WAY BY THE MANUFACTURER.