ELECSYS CMV IGG ASSAY
Received Sep 1, 2020 · Event occurred Jul 16, 2020
Report 1823260-2020-02147 · MDR key 10475868
Device
Generic name
Enzyme Linked Immunoabsorbent Assay, Cytomegalovirus
Manufacturer
Roche DiagnosticsModel number
NACatalog number
04784596190
Lot number
450074
Product problems
- False Negative Result
- False Negative Result
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
THE CUSTOMER'S CALIBRATION RECOVERY WAS FOUND TO BE ACCEPTABLE. THE CUSTOMER'S QC RECOVERY WAS FOUND TO BE ACCEPTABLE. THE PROVIDED SAMPLE WAS TESTED USING A RECOMLINE CMV IGG BLOT. THE BLOT TESTING FOUND THE SAMPLE TO SHOW NO CMV SPECIFIC REACTIVITY (NEGATIVE RESULT), WHICH CONFIRMS THE NEGATIVE RESULT THE CUSTOMER PREVIOUSLY RECEIVED ONSITE. THE INVESTIGATION FOUND THE CMV IGG ASSAY TO BE PERFORMING WITHIN SPECIFICATIONS.
Additional Manufacturer Narrative
THIS EVENT OCCURRED IN (B)(6). SAMPLE MATERIAL WAS REQUESTED FOR INVESTIGATION.
Description of Event or Problem
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR ELECSYS CMV IGG (CMV IGG) ON A COBAS 6000 E 601 MODULE. THE RESULT FROM THE E601 MODULE WAS < 0.150 U/ML (NEGATIVE) WITH A DATA FLAG. THE SAMPLE WAS REPEATED BY THE ABBOTT, LIAISON AND VIDAS METHODS WITH THE FOLLOWING RESULTS: ABBOTT: 14.7 U/ML (POSITIVE), LIAISON: 24 U/ML (POSITIVE), VIDAS: 8 U/ML (POSITIVE). ON (B)(6) 2020 THE RESULT FROM THE E601 MODULE WAS < 0.150 U/ML (NEGATIVE). THIS SAMPLE WAS ALSO REPEATED BY THE ABBOTT, LIAISON AND VIDAS METHODS WITH THE FOLLOWING RESULTS: ABBOTT: 17 U/ML (POSITIVE), LIAISON: 26 U/ML (POSITIVE), VIDAS: 6 U/ML (POSITIVE). THE ROCHE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE E601 MODULE SERIAL NUMBER WAS (B)(4).
Additional Manufacturer Narrative
THE SAMPLE WAS RECEIVED FOR INVESTIGATION AND TESTED WITH CMV IGG REAGENT 450074 ON AN E602 MODULE. THE CUSTOMER'S NEGATIVE CMV IGG RESULTS WERE REPRODUCED.