inforMED
MalfunctionFII

GAMBRO PHOENIX

Received May 12, 2008 · Event occurred Mar 17, 2008

Report 1047458 · MDR key 1047458

Device

Generic name

Hemodialysis Machine

Manufacturer

Gambro Inc.

Model number

PHOENIX

Catalog number

*

Lot number

*

Product problems

  • Alarm, audible
  • False Alarm
  • Repair

Patient

3 YR

  • No Patient Involvement

Narrative

Description of Event or Problem

AFTER PRESSING "PATIENT CONNECT" BUT BEFORE PATIENT WAS ACTUALLY CONNECTED, AIR IN BLOOD LINE ALARM OCCURRED. ATTEMPTED TO CLEAN VENOUS SENSOR BUT ALARM CONTINUED. TROUBLESHOOTING SHOWED POTENTIAL PROBLEM WITH ABD (AIR BUBBLE DETECTOR) MODULE, SO IT WAS REPLACED.