MalfunctionMZY
GRAFT,VASCULAR,STAINLESS STEEL TUNNELER
Received Aug 26, 2020 · Event occurred Jul 24, 2020
Report 10453679 · MDR key 10453679
Device
Generic name
Graft,vascular,stainless Steel Tunneler
Manufacturer
Geister Medizintechnik GmbhProduct problems
- Detachment of Device or Device Component
Patient
23725 DA
- Device Embedded In Tissue or Plaque
- Device Embedded In Tissue or Plaque
Narrative
Description of Event or Problem
PER MIDAS REPORT, BULLET ON THE WAREING TUNNELER DETACHED AFTER BEING INSERTED INTO PATIENT. AFTER INSPECTION, IT APPEARED ONE BALL BEARING THAT RETAINED THE BULLET TO THE TUNNELER WAS MISSING.
Description of Event or Problem
PER MIDAS REPORT, BULLET ON THE WAREING TUNNELER DETACHED AFTER BEING INSERTED INTO PATIENT. AFTER INSPECTION, IT APPEARED ONE BALL BEARING THAT RETAINED THE BULLET TO THE TUNNELER WAS MISSING.