inforMED
MalfunctionCDD

ELECSYS VITAMIN B12 II

Received Jul 31, 2020 · Event occurred Jul 6, 2020

Report 1823260-2020-01850 · MDR key 10352271

Device

Generic name

Radioassay, Vitamin B12

Manufacturer

Roche Diagnostics

Model number

B12 G2

Catalog number

07212771190

Lot number

44583501

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE CUSTOMER FOUND A BROWN MICROBEAD RING ON ONE OF THE AFFECTED REAGENT PACKS. BASED ON THE PROVIDED QC DATA, A GENERAL REAGENT ISSUE COULD MOST LIKELY BE EXCLUDED. THE INVESTIGATION IS ONGOING. THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem

THE INITIAL REPORTER RECEIVED QUESTIONABLE VITAMIN B12 G2 ELECSYS RESULTS FOR QC MATERIAL AND THREE PATIENT SAMPLES FROM COBAS E411 RACK ANALYZER SERIAL NUMBER (B)(4). THE CUSTOMER REPLACED THE REAGENT, RECALIBRATED THE ASSAY AND REPEATED THE SAMPLES. SAMPLE 1 INITIAL RESULT WAS 135.9 PG/ML AND THE REPEAT RESULT WAS 348.2 PG/ML. SAMPLE 2 INITIAL RESULT WAS 118.4 PG/ML AND THE REPEAT RESULT WAS 227.7 PG/ML. SAMPLE 3 INITIAL RESULT WAS 195.6 PG/ML AND THE REPEAT RESULT WAS 301.7 PG/ML. IT WAS UNKNOWN IF THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.

Additional Manufacturer Narrative

THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED. A HANDLING ISSUE AT THE CUSTOMER SITE COULD NOT BE EXCLUDED AS THE CUSTOMER USES THE REAGENT KITS LONGER THAN THE SPECIFIED STABILITY AS PER PRODUCT LABELING. INAPPROPRIATE STORAGE/TRANSPORTATION CONDITIONS AT THE CUSTOMER SITE COULD NOT BE EXCLUDED BASED ON THE BROWN MICROBEAD RING VISIBLE ON ONE OF THE AFFECTED KITS. A GENERAL REAGENT ISSUE CAN MOST LIKELY BE EXCLUDED AS THE CUSTOMER QC DATA WAS WITHIN THE SPECIFIED RANGES.