MOUTH PROP - SIL, BL, MED
Received Jul 15, 2020
Report 2523190-2020-00052 · MDR key 10280374
Device
Product problems
- Failure to Clean Adequately
- Failure to Clean Adequately
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
N/A.
Additional Manufacturer Narrative
COMPLAINT SAMPLE WAS NOT RETURNED TO MANUFACTURER. PER THE PROVIDED IMAGES, DARK RESIDUE HAD APPEARED WITHIN THE UPPER AND LOWER HOLES OF THE DEVICE. NO LOT NUMBER INFORMATION HAS BEEN PROVIDED AND MANUFACTURING RECORDS COULD NOT BE REVIEWED. THE COMPLAINT HAS BEEN CONFIRMED. VISIBLE RESIDUE BUILDUP ALONG THE METAL INSERTS IN THE UPPER AND LOWER HOLES. IMPROPER CLEANING PROCEDURE CAN LEAD TO BUILD UP OF RESIDUE. NO MANUFACTURING, WORKMANSHIP, OR MATERIAL DEFICIENCY HAS BEEN IDENTIFIED. NOTE THAT THE PRODUCT¿S LIFE EXPECTANCY IS 3 YEARS WHEN USED ONCE A DAY AND PRODUCT SHOULD BE INSPECTED PRIOR TO USE. PER THE IFU: STERILIZE INSTRUMENT / DEVICE ACCORDING TO AN APPROVED NATIONAL / INTERNATIONAL STANDARD. INTEGRA MILTEX INSTRUMENTS / PRODUCTS HAVE BEEN TESTED WITH NO ADVERSE EFFECTS UNDER THE FOLLOWING STERILIZATION PARAMETERS IN A STANDARD TABLE TOP STYLE STERILIZER: TEMPERATURE: 250ºF (121ºC) EXPOSURE TIME: 30 MINUTES. BETWEEN 05NOV2019 AND 30JUN2020, APPROXIMATELY 2,200 MDRS SUBMITTED ELECTRONICALLY BY INTEGRA LIFESCIENCES VIA TRACKWISE, INTEGRA'S COMPLAINT HANDLING SYSTEM, WERE NOT RECEIVED BY CDRH DUE TO A COMPUTER SYSTEM ISSUE. WITHIN THIS TIME PERIOD, AN ERROR WITH INTEGRA'S MIDDLEWARE, WHICH FACILITATES COMMUNICATIONS BETWEEN TRACKWISE AND THE FDA SYSTEM, CAUSED THE COMPLAINT RECORDS TO CLOSE AND INDICATE WE HAD RECEIVED AN ACKNOWLEDGEMENT 3 FROM THE FDA WHEN WE HAD NOT. INTEGRA INTERPRETED THE ACKNOWLEDGEMENT AS A SUCCESSFUL SUBMISSION; HOWEVER, SUBSEQUENT INVESTIGATION REVEALED THE ACKNOWLEDGEMENT 3 RECEIVED WAS FROM OUR MIDDLEWARE AND NOT FROM THE FDA (THESE ACKNOWLEDGEMENTS HAVE BEEN RETAINED AS PART OF THE DOCUMENTATION OF THE MDR). THE MALFUNCTION WAS RELATED TO THE RELOCATION OF THE TRACKWISE APPLICATION TO A NEW DATA CENTER DURING THE TRANSITION OF INTEGRA'S CORPORATE HEADQUARTERS FROM PLAINSBORO, NJ TO PRINCETON, NJ. PREVIOUSLY, INTEGRA HAD BEEN SUCCESSFULLY RECEIVING ACKNOWLEDGEMENTS 1, 2, AND 3 FROM THE FDA, AND OUR RECORDS REFLECT THESE ACKNOWLEDGEMENTS, INCLUDING THE DATE AND TIME STAMPS. CAPA: PR 229048 AND NC 20-011 HAVE BEEN OPENED BY INTEGRA TO FURTHER INVESTIGATE THE NONCONFORMANCE AND DEVELOP A CORRECTIVE ACTION PLAN. THE MIDDLEWARE ERROR HAS BEEN CORRECTED, AND INTEGRA HAS FILED MDRS SINCE THE CORRECTION AND VERIFIED THAT THE APPROPRIATE ACKNOWLEDGEMENTS HAVE BEEN RECEIVED FROM THE FDA. INTEGRA IS RESUBMITTING ALL IMPACTED MDR REPORTS FOR THE TIME PERIOD 05NOV2019 THROUGH 30JUN2020. INTEGRA LIFESCIENCES CONTACTED (B)(6), DIRECTOR OF REGULATORY PROGRAMS, OFFICE OF PRODUCT EVALUATION AND QUALITY AND (B)(6), ASSISTANT DIRECTOR, MDR TEAM, OFFICE OF PRODUCT EVALUATION AND QUALITY ON JULY 8-9, 2020 TO REPORT THESE ISSUES REGARDING MDR REPORTS.
Additional Manufacturer Narrative
THE DEVICE WAS NOT YET RECEIVED BY THE MANUFACTURER FOR EVALUATION. THE PLANT INVESTIGATION IS IN PROGRESS AND A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION. BETWEEN 05NOV2019 AND 30JUN2020, APPROXIMATELY 2,200 MDRS SUBMITTED ELECTRONICALLY BY INTEGRA LIFESCIENCES VIA TRACKWISE, INTEGRA'S COMPLAINT HANDLING SYSTEM, WERE NOT RECEIVED BY CDRH DUE TO A COMPUTER SYSTEM ISSUE. WITHIN THIS TIME PERIOD, AN ERROR WITH INTEGRA'S MIDDLEWARE, WHICH FACILITATES COMMUNICATIONS BETWEEN TRACKWISE AND THE FDA SYSTEM, CAUSED THE COMPLAINT RECORDS TO CLOSE AND INDICATE WE HAD RECEIVED AN ACKNOWLEDGEMENT 3 FROM THE FDA WHEN WE HAD NOT. INTEGRA INTERPRETED THE ACKNOWLEDGEMENT AS A SUCCESSFUL SUBMISSION; HOWEVER, SUBSEQUENT INVESTIGATION REVEALED THE ACKNOWLEDGEMENT 3 RECEIVED WAS FROM OUR MIDDLEWARE AND NOT FROM THE FDA (THESE ACKNOWLEDGEMENTS HAVE BEEN RETAINED AS PART OF THE DOCUMENTATION OF THE MDR). THE MALFUNCTION WAS RELATED TO THE RELOCATION OF THE TRACKWISE APPLICATION TO A NEW DATA CENTER DURING THE TRANSITION OF INTEGRA'S CORPORATE HEADQUARTERS FROM (B)(4). PREVIOUSLY, INTEGRA HAD BEEN SUCCESSFULLY RECEIVING ACKNOWLEDGEMENTS 1, 2, AND 3 FROM THE FDA, AND OUR RECORDS REFLECT THESE ACKNOWLEDGEMENTS, INCLUDING THE DATE AND TIME STAMPS. CAPA (B)(4) HAVE BEEN OPENED BY INTEGRA TO FURTHER INVESTIGATE THE NONCONFORMANCE AND DEVELOP A CORRECTIVE ACTION PLAN. THE MIDDLEWARE ERROR HAS BEEN CORRECTED, AND INTEGRA HAS FILED MDRS SINCE THE CORRECTION AND VERIFIED THAT THE APPROPRIATE ACKNOWLEDGEMENTS HAVE BEEN RECEIVED FROM THE FDA. INTEGRA IS RESUBMITTING ALL IMPACTED MDR REPORTS FOR THE TIME PERIOD 05NOV2019 THROUGH 30JUN2020. INTEGRA LIFESCIENCES CONTACTED (B)(6) ON JULY 8-9, 2020 TO REPORT THESE ISSUES REGARDING MDR REPORTS.
Description of Event or Problem
THE CUSTOMER REPORTED THAT AFTER CLEANING OF THE DLX114 MOUTH PROP - SIL, BL, MED DEVICE, A RESIDUE BUILD UP BETWEEN THE METAL INSERT AND THE SILICONE WAS FOUND. THERE WAS NO REPORTED PATIENT INCIDENT/INJURY.