inforMED
MalfunctionPMT

ELECSYS BRAHMS PCT

Received Jul 13, 2020 · Event occurred Jun 1, 2020

Report 1823260-2020-01671 · MDR key 10264658

Device

Generic name

Device To Detect And Measure Procalcitonin (pct) In Human Clinical Specimens

Manufacturer

Roche Diagnostics

Model number

PCT

Catalog number

08828679200

Lot number

466361

Product problems

  • Unexpected Therapeutic Results
  • Unexpected Therapeutic Results

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED. THE CALIBRATION AND QC RESULTS DID NOT INDICATE ANY ISSUE WITH THE REAGENT OR ANALYZER.

Additional Manufacturer Narrative

THE INVESTIGATION IS ONGOING. THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem

THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS BRAHMS PCT RESULTS FOR ONE PATIENT FROM COBAS E 801 MODULE SERIAL NUMBER (B)(4) WHICH DO NOT FIT INTO CLINICAL PICTURE. THE PATIENT RECEIVES HEMODIALYSIS AND WAS TREATED SUCCESSFULLY WITH ANTIBIOTICS 2 WEEKS AGO. THE CRP RESULTS CONSTANTLY DECREASED (CURRENTLY AROUND 30 MG/DL), BUT PCT REMAINS ELEVATED AT 100 NG/ML ALL THE TIME. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY AND WERE QUESTIONED BY THE PHYSICIAN.